background image

4

Signet 8250-2 Level Transmitter 

Zero reference point (Z):

The point in the vessel where you want the 8250 to display zero 
(0 ft., 0 gal. etc).

•  If Z is located below the 

fl

 uid surface, the 8250 will display 

a positive level measurement.

•  If Z is located above the 

fl

 uid surface, the 8250 will display 

a negative level measurement. 

Sensor Location point (S

Loc

):

The point on the level sensor where the measurement is taken.

•  The 2450 Pressure sensor measures from the centerline of 

the diaphragm.

Offset (O):

The distance from Z to S

Loc

.

•  Enter a positive value in the Calibrate menu if the sensor is 

located above Z.

•  Enter a negative value in the Calibrate menu if the sensor 

is located below Z.

•  Enter 0 in the Calibrate menu if the sensor is located at Z.

Level (L):

The distance from Z to surface of 

fl

 uid (displayed as "Level"

by transmitter).

L

 > 0

O

 > 0

Z

S

LOC

L

 > 0

O

 < 0

Z

S

LOC

< 0

S

LOC

O

 < 0

Z

2450 Pressure Sensor Location and Offset

Z

Z

For horizontal cylinders only, the 
zero reference point MUST be 
the lowest point in the vessel.

For most vessels, the zero 
reference point may be 
designated as any height 
in the vessel.

5. System 

Confi guration

The 8250 Level Transmitter must be programmed using the following system con

fi

 guration procedure.

1.  Determine where you want the level measurement to start. This is the Zero reference point (

Z

).

 

Review the diagram to help select the best option.

2.  Determine where you will mount the sensor. This is 

S

Loc

.

 

Consult the Sensor manual for information regarding the best location for the sensor.

3.  Measure the distance between Z and S

Loc

. This is 

O(ffset).

4.  Enter the Offset into the Calibrate menu.

Summary of Contents for Signet 3-8250-2

Page 1: ...ble and reversible Max loop impedance 50 max 12 V 325 max 18 V 600 max 24 V Update rate 300 ms Output accuracy 0 03 mA Relay outputs 2 sets mechanical SPDT contacts Software supports 2 additional rela...

Page 2: ...ure instrument temporarily with tape from front or grip from rear of instrument DO NOT RELEASE 5 Press quick clips outward and remove 2 1 Terminals 11 13 Digital S3 L The 2450 sensor and External Rela...

Page 3: ...el 1 4 to 20 mA in Channel 2 4 to 20 mA in System Pwr Loop System Pwr Loop AUX Power AUX Power System Pwr Loop System Pwr Loop AUX Power AUX Power Loop Input 4 to 20 mA System Pwr Loop System Pwr Loop...

Page 4: ...menu if the sensor is located at Z Level L The distance from Z to surface of fluid displayed as Level by transmitter L 0 O 0 Z SLOC L 0 O 0 Z SLOC L 0 SLOC O 0 Z 2450 Pressure Sensor Location and Offs...

Page 5: ...ted at each transition point in the vessel shape for example at custom point 9 where the shape changes from a cylinder to a cone The more complex sections should be defined with more points Notice tha...

Page 6: ...to change display selection Output settings cannot be edited from the VIEW menu When using the CALIBRATE or OPTIONS menus the ProcessPro will return to the VIEW menu if no key is pressed for 10 minut...

Page 7: ...g Procedure Step 1 Press and hold ENTER key 2 seconds to select the CALIBRATE menu 5 seconds to select the OPTIONS menu Step 2 The Key Code is UP UP UP DOWN keys in sequence After entering the Key Cod...

Page 8: ...L Single point calibration for Level Reset LVL to Factory Cal Yes Reset Level to original factory calibration Set LVL entry is erased No Return to menu do not reset Level Loop Source LVL Select the so...

Page 9: ...for relay response Relay will be activated only if the source value exceeds the setpoint for this time period Relay settings will repeat for Relay 2 and for External Relays External Relays require th...

Page 10: ...t Test Relay 1 Press UP or DOWN keys to manually toggle the state of Relay 1 Display Factory settings shown Description VOL Shape Vert Cylinder Select the shape of the vessel where Sensor is located V...

Page 11: ...iew all custom points and reset values so each point is greater than the preceding point CHK SENSOR Problem with the sensor during different modes of operation 1 Wiring problem AUXILIARY POWER was not...

Page 12: ...upply Note When using the 8059 4 or 8059 4AC only relays 1 and 2 can be utilized with the flow instrument Accessories Mfr Part No Code Description 3 8050 159 000 184 Universal mounting kit 3 8050 395...

Reviews: