8

Signet 8250-2 Level Transmitter 

8. Calibrate 

Menu

Display

(Factory settings shown)

Description

LVL Units:

ft >

Select the units of measure: ft, in, m,cm.

Show LVL in %:

No >

Yes: Level will be displayed as a percentage of the full scale.
No: Level will be displayed as engineering units.

LVL 100%:

+0.0000 ft

 

>

If "Show LVL in %" is YES, enter the full scale value. The units are the same as LVL Units.

Sensor Offset:

+0.0000 ft 

>

Enter the distance from S

LOC

 to Z in the vessel.

Enter 

+

 if the sensor is located above Z.

Enter 

-

 if the sensor is located below Z. 

VOL Units:

gal >

Select the units of measure for Volume: gal, ft

3

, in

3

, m

3

, kg, lb, or lit (liters)

Show VOL in %:

No >

Yes: Volume will be displayed as a percentage of the full scale.
No: Volume will be displayed as engineering units.

VOL 100%:

+0.0000 gal 

>

If "Show VOL in %" is YES, enter the full scale value. The units are the same as VOL Units

Spec Gravity:

1.0000 >

Enter the speci

fi

 c gravity of the 

fl

 uid. Factory setting is 1 (water)

This setting is required only if the level measurement is made by a pressure sensor
(Signet 2450), or if volume units are selected ( kg or lb )

Set LVL:

 

>

Single-point calibration for Level.

Reset LVL

to Factory Cal: 

>

Yes: Reset Level to original factory calibration. ("Set LVL" entry is erased.)
No: Return to menu, do not reset Level.

Loop Source:

LVL >

Select the source for the 4-20 mA Loop: LVL or VOL.

Loop Rng: ft

+0.0000 

 +10.000  >

Select the minimum and maximum values for the 4-20 mA Current loop output.

Relay 1 Mode:

Off >

Select the mode of operation for Relay Output: Low, High, or Window.
The relay may be disabled (Off) if not in use.

Reset Sensor

Con

fi

 guration: >

Use this function during initial startup of a new system, and whenever a new sensor is added 
to the system. It enables the 8250 to scan the sensor input terminals and determine the type 
and speci

fi

 c serial numbers of the sensors connected to the sensor input.

Technical Note

If you see this error

The most likely cause is

Remedy

Too Much Error

Check Sensor

A new sensor was connected to the 8250 
but the "Sensor Reset Con

fi

 guration" has 

not been executed.

Perform the sensor con

fi

 guration.

*

*

IMPORTANT TECHNICAL ADVICE. DO THIS FIRST!

Before proceeding with the calibration, press the UP key twice, then press the RIGHT key to restore the function of the con

fi

 guration of 

the sensor (sensor con

fi

 guration RESET). This procedure is part of the sensor 8250 and provides for many calibration functions.

Summary of Contents for Signet 3-8250-2

Page 1: ...ble and reversible Max loop impedance 50 max 12 V 325 max 18 V 600 max 24 V Update rate 300 ms Output accuracy 0 03 mA Relay outputs 2 sets mechanical SPDT contacts Software supports 2 additional rela...

Page 2: ...ure instrument temporarily with tape from front or grip from rear of instrument DO NOT RELEASE 5 Press quick clips outward and remove 2 1 Terminals 11 13 Digital S3 L The 2450 sensor and External Rela...

Page 3: ...el 1 4 to 20 mA in Channel 2 4 to 20 mA in System Pwr Loop System Pwr Loop AUX Power AUX Power System Pwr Loop System Pwr Loop AUX Power AUX Power Loop Input 4 to 20 mA System Pwr Loop System Pwr Loop...

Page 4: ...menu if the sensor is located at Z Level L The distance from Z to surface of fluid displayed as Level by transmitter L 0 O 0 Z SLOC L 0 O 0 Z SLOC L 0 SLOC O 0 Z 2450 Pressure Sensor Location and Offs...

Page 5: ...ted at each transition point in the vessel shape for example at custom point 9 where the shape changes from a cylinder to a cone The more complex sections should be defined with more points Notice tha...

Page 6: ...to change display selection Output settings cannot be edited from the VIEW menu When using the CALIBRATE or OPTIONS menus the ProcessPro will return to the VIEW menu if no key is pressed for 10 minut...

Page 7: ...g Procedure Step 1 Press and hold ENTER key 2 seconds to select the CALIBRATE menu 5 seconds to select the OPTIONS menu Step 2 The Key Code is UP UP UP DOWN keys in sequence After entering the Key Cod...

Page 8: ...L Single point calibration for Level Reset LVL to Factory Cal Yes Reset Level to original factory calibration Set LVL entry is erased No Return to menu do not reset Level Loop Source LVL Select the so...

Page 9: ...for relay response Relay will be activated only if the source value exceeds the setpoint for this time period Relay settings will repeat for Relay 2 and for External Relays External Relays require th...

Page 10: ...t Test Relay 1 Press UP or DOWN keys to manually toggle the state of Relay 1 Display Factory settings shown Description VOL Shape Vert Cylinder Select the shape of the vessel where Sensor is located V...

Page 11: ...iew all custom points and reset values so each point is greater than the preceding point CHK SENSOR Problem with the sensor during different modes of operation 1 Wiring problem AUXILIARY POWER was not...

Page 12: ...upply Note When using the 8059 4 or 8059 4AC only relays 1 and 2 can be utilized with the flow instrument Accessories Mfr Part No Code Description 3 8050 159 000 184 Universal mounting kit 3 8050 395...

Reviews: