GLX-RSS-2-300 WL User Manual
26
GLX-RSS-2-300 WL Non-Contact Flow Meter
Tilt angle report
$RDANG,A*CSUM<CR><LF>
$RDANG:
The keyword sent in the beginning of each tilt angle report.
A:
The measured tilt angle, in degrees, 0 being horizontal.
CSUM:
The check sum of the characters in the report from $ to * excluding these
characters.
Signal SNR report
$RDSNR,S1,S2*CSUM<CR><LF>
$RDSNR:
The keyword sent in the beginning of each SNR report.
S1:
The current SNR of measured velocity signal in dBm.
S2:
The average SNR of measured velocity signal in dBm.
CSUM:
The check sum of the characters in the report from $ to * excluding these
characters.
Quality of signal report
$QOS,Q1,Q2*CSUM<CR><LF>
$QOS:
The keyword sent in the beginning of each quality of signal report.
Q1:
Quality of signal estimate based on sensor vibration.
Q2:
Quality of signal estimate based on radar signal analysis.
CSUM:
The check sum of the characters in the report from $ to * excluding these
characters.
Quality of signal factor:
0 – Excellent measurement quality
1 – Good measurement quality
2 – Low measurement quality
3 – Unacceptable measurement quality
Discharge report
$DIS,D*CSUM<CR><LF>
$DIS:
The keyword sent in the beginning of each discharge report.
D:
The measured discharge, in defined units.
CSUM:
The check sum of the characters in the report from $ to * excluding these
characters.