GLX-RSS-2-300 WL Non-Contact Flow Meter
3. Check that the distance between the instrument and the water surface is
within the instrument’s operating range.
Use a tape measure or a similar tool to measure the distance between the
instrument’s antenna and the water surface. If the distance to the water surface is
greater than the instrument’s operational range, the instrument will not be able to
measure the distance and the water level.
4. Check that the distance between the instrument and the water surface is
within configured active zone.
Connect the instrument to the Geolux Instrument Configurator PC application
over RS-232 connection. Check the active zone parameters – if the distance to the
actual water surface is outside of the predefined active zone, the instrument will
not be able to report the correct distance. Modify the active zone parameters so
that the actual distance to the water surface is within the active zone.