PSY-201 User Guide
GP-UM-PSY-201-11
Page 97 of 105
Note that the “total time length” and “current time” boxes use the actual time stamps for
recorded data. Therefore, the speed setting only affects the update speed of the counter.
Markers
This option allows the user to mark points on the sphere and measure the angles between
marked points or between marked points and a reference SOP. Up to 99 points can be marked.
Selecting the Marker option from the Measurement menu causes the Marker control pane to
appear on the left side of the screen.
Reference: The top section lists the current reference SOP. This can be changed by
selecting Reference SOP from the Options menu.
Markers: The middle section lists the Stokes parameters of all marked points. The 4th
parameter position can be configured to display S0, DOP (%), Azimuth, Ellipticity,
DLP, DCP, dSOP (SOP difference between consecutive markers, in degrees), dDOP
(DOP difference between consecutive markers, in %), dPower (power difference
between consecutive markers, in dB), dREF (SOP difference between marker and
reference SOP, in degrees), dREF dop (DOP difference between marker and
reference), or dREF pow (power difference between marker and reference). If
dREF or dSOP are selected, PolaView will display vector lines from the origin to the
markers for visual clarity of the angles. If dREF is selected, the reference SOP will
also be shown as a green marker, with a vector line to the origin.