PSY-201 User Guide
GP-UM-PSY-201-11
Page 49 of 105
Polarization Scrambling
Type
Trace
Scrambling Rate Range
Triangle
Continuous
0.01 to 2000 ×2
π
/s
Discrete
Random, discontinuous points
0.01 to 20,000 points/s
Tornado Continuous,
spiral-type
pattern with rotating or fixed
spin axis
0.01 to 2000 Rev/s
Triggered
Triggered random state
generation
One point per rising TTL pulse
Polarization Modulation
Waveforms
Sine, square, triangle
Frequency
0.01 to 1000 Hz
Amplitude
Range (Radian)
Range (
π
) Range
(V)
4.71 1.5
45
Offset Depends
on
amplitude
The relationship between Amp and Ofs is:
Amp
≤
Ofs
Ofs range: A to (2A
max
– A),
where A
max
is the value listed in the amplitude table
Notes:
Unless otherwise noted, specifications listed in table apply for standard 1480-1620nm or 1280-
1340nm operation at 23±5°C, at power levels >
−
25 dBm.
1.
Optimal sampling rate ranges depend on data collection/display mode and setup. Some
examples:
Externally triggered measurement:
Sampling rate can be up to 4M points/s.
PolaView Poincaré sphere mode with trace sensitivity = 0: Sampling rate should be
≤
500K
points/s for best results.
2.
Bandwidth may change at lower power levels due to automatic gain control.
3.
Accuracy depends on input SOP variation speed. See SOP accuracy specification for input
SOP variation up to 10
π
/s.