PSY-201 User Guide
GP-UM-PSY-201-11
Page 37 of 105
In manual control mode, the user can set a fixed integration time. The minimum
integration time is 0.25
μ
s, which corresponds to the highest sampling bandwidth of the PSY-201
(4M SOP samples/second). Both the time value and unit can be changed.
Range: 0.25 µs to 180.00 s
Default: 1µs
After entering the integration time setup, use the
▲▼
buttons to select auto or manual
integration time control. For manual control, use the
◄►
buttons to move the cursor position to
time value or time units, and the
▲▼
buttons to change the value of the selected digit or scroll
through unit options.
When finished with integration time setup, press ENTER. If the setting was changed from
the previous setting, an “OK” will appear on the right side of the screen to confirm the change.
Analog Outputs
The PSY-201 can output up to 4 monitor voltages that can be configured to monitor
different parameters. The DB15 expansion port on the rear panel is used for the monitor voltages.
See Figure 7 for the connector pinout.
To configure the analog outputs, use the
◄►
buttons to move the cursor position and the
▲▼
buttons to change the value of the selected digit or parameter.
Aout = monitor voltage index. Range: 1 to 4.
Monitor parameter options: power, S1, S2, S3, DOP, dREF
Vlim = upper limit of output voltage range. Range: 0 to 5V.
Min = monitored parameter value that maps to 0V.
Max = monitored parameter value that maps to Vlim.
Ranges for parameter limits:
Power:
−
60 to +15 dBm (Note: limit setup is in dBm, but monitor voltage
still scales linearly with power)
S1/S2/S3:
−
1 to +1
DOP:
0
to
200%
dREF:
0 to 180.0° (Note: degrees only; cannot be set to radians)
Aout
1 : Power
Vlim
5.0
Min:
-60.0
Max:
+15.0
SET
L
V
dBm
[Set]
Unit indicator
S1, S2, S3: unitless
DOP: %
Power: dBm
dREF: deg
Integration Time
Manual: 000.50
us
SET
OK
L