5.4. Adjusting the print mode
This function sends the measurement data to a printer or PC.
A. Printer:
In this mode, measurement data in a report format will be sent, as stated below, through
the Serial Port RS232.
If the G650
i
meter has a built-in Thermal Printer, this report will be printed and
simultaneously sent through the Serial Port RS232.
B. Computer:
The Computer Mode just sends the data, without considering formats. The data is
separated by “;” in order to make the separation easier. This report may be easily captured
and interpreted by a computerized system connected to the RS232 Serial Port.