38
I
Moisture Meter G650
i
If the error persists, there was a failure in the instrument; send the instrument to Gehaka
Service Department.
Error 7 - Delta Temp. > 15 ºC
This error indicates that the temperature difference between the instrument and the
sample is greater than 15 ºC. Wait for some minutes with the sample in the G650
i
drawer
so this difference stays below 15 ºC. It is recommended that the sample temperature and
instrument temperature shall be as close as possible, whenever possible.
If the error persists, there was a failure in the instrument; send the instrument to Gehaka
Service Department.
Error 8 – Low Sample Weight!
Error 9 – High Sample Weight!
After having the sample weight within the limits, a final verification is made after loading;
if the amount of sample used is outside the limits that G650
i
is able to correct, this sample
will be unloaded. Increase or reduce the amount of sample to correct the error.
If the error persists, there was a failure in the instrument; send the instrument to Gehaka
Service Department.
Erro 10 - Umid. Amostra Baixa!
Erro 11 - Umid. Amostra Alta!
The G650
i
was calibrated using the standard oven method, and its precision is ensured in
the range where the tests and the adjustment were performed. When the measurement
is out of this range, the G650
i
will display this message. If the measurement is actually
out of the scale range, we recommend creating a new Equation using the resources in the
website
www.moisturetester.com.br
. See item “5.8 Uploading Internet equation” for more
details of the procedure.
Error 12 - Access Locked!
This message is displayed when we attempt to enter the “Configuration” function without
the Hard Lock key that releases the access to that function. This key shall be installed in the
G650
i
serial output. It is the security resource to prevent G650
i
parameters from being
inadvertently changed.