28
I
Moisture Meter G650
i
Makes a self-diagnosis of G650
i
, and issues a report with all data. The report will be sent to
serial output RS232 or to the thermal printer connected to G650
i
serial port.
To get the report:
1.
Connect the G650
i
;
2.
Go to “Configuration” and depress YES; do not forget to put the Hard Lock in the
serial output;
3.
Go to “5.9 Self-report” function as depress YES;
4.
On the screen several information will show up related to G650
i
self-report, at the
end we will have the message below stating the report printing:
5.
Wait for the report to be printed;
This report provides the Service with data for assessing whether the G650
i
is operating
properly, after having passed through the self-diagnosis when it is connected.
Note that there are two groups of information, the first one indicates the G650
i
operation,
the second one the Products that are available in G650
i
with the Calibration Curve Version
and the minimum and maximum moisture values.
Wait, issuing
Report...
5.9. Self-report