XMTC User’s Manual
49
Chapter 5. Specifications
Chapter 5. Specifications
5.1 Performance
Accuracy
±2% of span
Linearity
±1% of span
Repeatability
±0.5% of span
Stability
Zero:
±0.5% of span per week
Span
: ±0.5% of span per week
Response Time
20 seconds for 90% of step change
Measurement Ranges
(typical)
0-2%
0-5%
0-10%
0-25%
0-50%
0-100%
50-100%
80-100%
90-100%
95-100%
98-100%
Measurement Gases
(typical)
H
2
in N
2,
air or CO
2
He in N
2
or Air
CO
2
in N
2
or Air (minimum range 0-20% CO
2
)
SO
2
in Air (minimum range 0-10% SO
2
)
Ar in N
2
or air (minimum range 0-20% argon)
Ambient Temperature Effect
±0.05% of span per °C
Required Sample Gas
Flow Rate
0.1 to 4.0 SCFH (5 to 2000 cc/min), 0.5 SCFH (250 cc/min) nominal
Required Reference Gas
Flow Rate
0.01 to 4.0 SCFH (5 to 2,000 cc/min), 0.5 SCFH (250 cc/min) nominal
Summary of Contents for XMTC
Page 1: ...GE Measurement Control Gas Analysis 910 217 Rev G October 2013 XMTC User s Manual ...
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 10: ...Contents viii XMTC User s Manual no content intended for this page ...
Page 18: ...Chapter 1 Features and Capabilities 8 XMTC User s Manual no content intended for this page ...
Page 58: ...Chapter 4 Calibration 48 XMTC User s Manual no content intended for this page ...
Page 62: ...Chapter 5 Specifications 52 XMTC User s Manual no content intended for this page ...
Page 68: ...Appendix A Supplemental Information 58 XMTC User s Manual no content intended for this page ...
Page 80: ...70 XMTC User s Manual no content intended for this page ...
Page 118: ...Index 108 XMTC User s Manual no content intended for this page ...
Page 120: ...Warranty 110 XMTC User s Manual no content intended for this page ...
Page 124: ... no content intended for this page ...
Page 125: ......