Appendix B. Typical Applications
B.1.4 Permanent Installation
Continuous monitoring of the furnace atmosphere H
2
content using the XMTC assures a high degree of quality control
in the manufacturing process.
B.1.5 Specifications
Typical Ranges:
0 to 10% H
2
in N
2
0 to 25% H
2
in N
2
0 to 100% H
2
in N
2
Operating Conditions:
Pressure:
Ambient
Temperature:
+540 to +1370
o
C (+1000 to +2500
o
F)
B.1.6 Detailed Operating Procedure
The following procedure details the start-up, operation, and calibration of the 2-port (Sealed Reference Gas) XMTC
sample system for heat treat furnace applications shown in
Figure 56 on page 59
.
Needle valves N1 through N4 on the sample system drawing have the following functions:
•
N1 – selects/isolates the process sample gas
•
N2 – selects/isolates the calibration zero gas
•
N3 – selects/isolates the calibration span gas
•
N4 – controls the flow of the selected gas
Summary of Contents for XMTC
Page 1: ...GE Measurement Control Gas Analysis 910 217 Rev G October 2013 XMTC User s Manual ...
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 10: ...Contents viii XMTC User s Manual no content intended for this page ...
Page 18: ...Chapter 1 Features and Capabilities 8 XMTC User s Manual no content intended for this page ...
Page 58: ...Chapter 4 Calibration 48 XMTC User s Manual no content intended for this page ...
Page 62: ...Chapter 5 Specifications 52 XMTC User s Manual no content intended for this page ...
Page 68: ...Appendix A Supplemental Information 58 XMTC User s Manual no content intended for this page ...
Page 80: ...70 XMTC User s Manual no content intended for this page ...
Page 118: ...Index 108 XMTC User s Manual no content intended for this page ...
Page 120: ...Warranty 110 XMTC User s Manual no content intended for this page ...
Page 124: ... no content intended for this page ...
Page 125: ......