Appendix B. Typical Applications
66
XMTC User’s Manual
B.2.4 How Previously Handled
The system (generator) was leak checked periodically. If a leak occurred between checks, an explosion could occur.
Moisture analyzers were also used for continuous analysis, since the presence of moisture in the H
2
is indirect evidence
of an air leak.
B.2.5 Permanent Installation
Continuous monitoring of the generator H
2
purity using the XMTC provides increased safety. A low H
2
reading alerts
plant personnel to a potential safety problem and allows them to locate the leak and correct the problem.
B.2.6 Specifications
Range:
80 to 100% H
2
in N
2
Operating Conditions:
Pressure:
0.5 to 75 psig
Temperature:
+30° to +50
o
C (+86 to +122
o
F)
B.2.7 Detailed Operating Procedure
The following procedure details the start-up, operation, and calibration of the 4-port (Flowing Reference Gas) XMTC
sample system for the hydrogen purity applications shown in
Figure 58 on page 64
.
Needle valves N1 through N4 on the sample system drawing have the following functions:
•
N1 – controls the flow of the process sample gas
•
N2 – controls the flow of the zero calibration gas
•
N3 – controls the flow of the span calibration gas
•
N4 – controls the flow of the reference gas
Summary of Contents for XMTC
Page 1: ...GE Measurement Control Gas Analysis 910 217 Rev G October 2013 XMTC User s Manual ...
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 10: ...Contents viii XMTC User s Manual no content intended for this page ...
Page 18: ...Chapter 1 Features and Capabilities 8 XMTC User s Manual no content intended for this page ...
Page 58: ...Chapter 4 Calibration 48 XMTC User s Manual no content intended for this page ...
Page 62: ...Chapter 5 Specifications 52 XMTC User s Manual no content intended for this page ...
Page 68: ...Appendix A Supplemental Information 58 XMTC User s Manual no content intended for this page ...
Page 80: ...70 XMTC User s Manual no content intended for this page ...
Page 118: ...Index 108 XMTC User s Manual no content intended for this page ...
Page 120: ...Warranty 110 XMTC User s Manual no content intended for this page ...
Page 124: ... no content intended for this page ...
Page 125: ......