Chapter 3. Operation and Programming
30
XMTC User’s Manual
3.7.2 4 and 20 mA Cal (cont.)
Figure 29:
Numeric Calibration
Window
Enter the desired number in the text box and click on
Next Item/Enter.
(Click on
Previous Item
or
Exit Page
to close the
window without changing the signal.) You are then returned to the previous window. After you have calibrated the
desired input signal, click on
STORE
to save the entry, and on
Next Item/Enter
to confirm it. If the signal is not
satisfactory, click on
ABORT
. (Click
Previous Item
or
Exit Page
to close the window without changing the signal.) You
are then returned to the
4-20 mA Output
window (see
Figure 25 on page 28
).
3.7.3 4-20 mA % Test
The window for
4-20 mA % Test
is similar to
Figure 30
below.
Figure 30:
4-20 mA% Test
Window
Enter the desired percentage, and click on
Next Item/Enter
to confirm the entry. A second text box opens, enabling you
to test at another percentage if desired. Repeat the procedure until you have entered all desired test values. Then click
Exit Page
to close the window.
Summary of Contents for XMTC
Page 1: ...GE Measurement Control Gas Analysis 910 217 Rev G October 2013 XMTC User s Manual ...
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 10: ...Contents viii XMTC User s Manual no content intended for this page ...
Page 18: ...Chapter 1 Features and Capabilities 8 XMTC User s Manual no content intended for this page ...
Page 58: ...Chapter 4 Calibration 48 XMTC User s Manual no content intended for this page ...
Page 62: ...Chapter 5 Specifications 52 XMTC User s Manual no content intended for this page ...
Page 68: ...Appendix A Supplemental Information 58 XMTC User s Manual no content intended for this page ...
Page 80: ...70 XMTC User s Manual no content intended for this page ...
Page 118: ...Index 108 XMTC User s Manual no content intended for this page ...
Page 120: ...Warranty 110 XMTC User s Manual no content intended for this page ...
Page 124: ... no content intended for this page ...
Page 125: ......