Chapter 4. Calibration
44
XMTC User’s Manual
4.3 Required Equipment and Materials
To calibrate the XMTC, you will need the following equipment and materials:
•
XMTC Calibration Sheet
•
+24 VDC, 1.2 A power supply (system power, if unit is installed in a system)
•
Zero gas
•
Span gas
•
Reference gas (for the 4-Port version - the reference gas is usually the same as the span gas.)
•
Sample system or individual components (e.g. flow meter, needle valve, pressure regulator) for connecting
zero and span gases and controlling pressure and flow rates
Note:
The accuracy of the calibration will only be as accurate as the composition accuracy of the zero and span
gases.
WARNING!
The calibration procedure described in this section requires the use of specialized
apparatus and should be performed only by properly trained personnel.
WARNING!
Do not use explosive gas mixtures to calibrate the XMTC.
Summary of Contents for XMTC
Page 1: ...GE Measurement Control Gas Analysis 910 217 Rev G October 2013 XMTC User s Manual ...
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 10: ...Contents viii XMTC User s Manual no content intended for this page ...
Page 18: ...Chapter 1 Features and Capabilities 8 XMTC User s Manual no content intended for this page ...
Page 58: ...Chapter 4 Calibration 48 XMTC User s Manual no content intended for this page ...
Page 62: ...Chapter 5 Specifications 52 XMTC User s Manual no content intended for this page ...
Page 68: ...Appendix A Supplemental Information 58 XMTC User s Manual no content intended for this page ...
Page 80: ...70 XMTC User s Manual no content intended for this page ...
Page 118: ...Index 108 XMTC User s Manual no content intended for this page ...
Page 120: ...Warranty 110 XMTC User s Manual no content intended for this page ...
Page 124: ... no content intended for this page ...
Page 125: ......