Using Vscan Air CL
5-86
Vscan Air
–
User Manual
Direction GP092020-1EN
Rev 18
Assessing Display Device Wi-Fi Performance with Vscan Air probe
Prior to setting up for a guided procedure, it is recommended to
check if the display device being planned for use during the
procedure supports a stable connection with the Vscan Air
probe, by following the below steps.
If the display device does not pass the above criteria
successfully, using an alternate display device is recommended
for supporting guided procedures.
Step
Step name
Description
1
Configure app settings
Set the Video duration to 1 second from Menu -> Configuration.
This will help to visualize discontinuities in loop playback which
indicate lost frames.
2
Connect probe and app
Connect the probe with the app
3
Select preset with color
Select Small Parts from the Shallow (Linear) presets menu.
Enter color flow mode by pressing the ROI button. Adjust the
color gain to 100% to visualize noise inside the color ROI.
4
Confirm/Verify
Observe the noise pattern inside the color ROI. The randomness
of the pattern should not appear to pause occasionally.
5
Review cine
Press freeze. Then press the Play button to the left of the scroll
bar.
6
Confirm/Verify
Observe the blue cine position indicator as it moves from left to
right along the position indicator. It should move in a consistent
fashion from left to right (the distance it moves with each
increment appears equal).
Summary of Contents for Vscan Air
Page 5: ...Vscan Air User Manual i 3 Direction GP092020 1EN Rev 18 This page intentionally left blank...
Page 14: ...i 12 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 324: ...Using Vscan Air CL 5 112 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 370: ...Vscan Air Maintenance 6 46 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 382: ...Privacy and Security 7 12 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 406: ...Index 2 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 407: ......