Personal Information Collected by the Product
Vscan Air
–
User Manual
7-11
Direction GP092020-1EN
Rev 18
Potential Hazardous Situations Resulting From Failures of the IT
Network
Hazardous situations
The following situations have been identified as potentially
hazardous as a result of the IT network failing to provide the
required characteristics specified above.
1. Delayed or impaired access to images or other exam
information or patient data.
2. Permanent loss of images or other exam information or
patient data.
3. Corruption of images or other exam information or patient
data.
Warning
In addition to the hazardous situations identified above,
connection of the Vscan Air to a network that includes other
equipment could result in other unidentified risks to patients,
operators or third parties. The responsible organization should
identify, analyze, evaluate and control these risks on an ongoing
basis. Re-evaluate the risks if any of the following occurs:
1. Changes in network configuration
2. Connection of additional items to the network
3. Disconnecting items from the network
4. Update of equipment connected to the network
5. Upgrade of equipment connected to the network
Summary of Contents for Vscan Air
Page 5: ...Vscan Air User Manual i 3 Direction GP092020 1EN Rev 18 This page intentionally left blank...
Page 14: ...i 12 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 324: ...Using Vscan Air CL 5 112 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 370: ...Vscan Air Maintenance 6 46 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 382: ...Privacy and Security 7 12 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 406: ...Index 2 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 407: ......