Configuration
Vscan Air
–
User Manual
4-5
Direction GP092020-1EN
Rev 18
3.
Time Gain Compensation (TGC) Control
(see
) - Tap the
ON/OFF
toggle switch
to turn ON or OFF TGC controls. When TGC is ON, you can
adjust the black and white image via the TGC controls on
the scan screen.
NOTE:
For adjustment up to 6 TGC controls are available in portrait
and landscape layouts.
NOTE:
The active gain control is disabled as long as the TGC
controls are active.
4.
Cardiac Flip L/R (see
)-
Setting the
Cardiac Flip L/R
option true, will flip the cardiac
image and orientation marker horizontally to right and vice
versa.
•
Flip Left Right option is available for all the presets and
it will have impact only when Cardiac preset is selected.
The default value of this configuration is false which
means it is flipped left.
Figure 4-4. CLA-Cardiac (no flip)