Cleaning using a cleaning solution
1. Prepare a basin with enzymatic cleaner per the
manufacturer’s instructions.
2. Immerse probe in the cleaning solution and ensure no air
bubbles are trapped.
3. Use a soft nylon bristle brush to clean the probe. Do Not use
the brush on the probe lenses as this can damage the soft
acoustic lens.
4. Ensure the probe remains in the cleaning solution for the
minimum contact time listed on the enzymatic cleaner label.
Make sure to follow manufacturers’ recommendations.
5. Visually inspect the probe and probe components for soil.
Repeat steps 4 - 6 until all visible soil has been removed
from the surface of the probe.
6. Thoroughly rinse the probe to remove traces of the cleaner
solution. Follow the cleaner manufacturers’ instructions.
7. Visually inspect the device in a well-lit area to ensure all
surfaces are free from residual cleaning solution. Repeat
Step 6 if visible cleaning solution is observed.
8. Thoroughly pat dry the probe using a clean low-lint soft, dry
disposable cloth or wipe. Do not use abrasive paper
products.
Summary of Contents for Vscan Air
Page 5: ...Vscan Air User Manual i 3 Direction GP092020 1EN Rev 18 This page intentionally left blank...
Page 14: ...i 12 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 324: ...Using Vscan Air CL 5 112 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 370: ...Vscan Air Maintenance 6 46 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 382: ...Privacy and Security 7 12 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 406: ...Index 2 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 407: ......