GE
DRAFT
V
OLUSON
™ P8/V
OLUSON
™P6
DIRECTION 5775469, R
EVISION
3
DRAFT (J
ULY
19, 2018)
B
ASIC
S
ERVICE
M
ANUAL
Chapter 5 - Components and Functions (Theory)
5-57
5-14-3-8
Export System Data
Select the EXPORT SYSTEM DATA button on the “Service Tools” page to Full Backup the System
State. This includes dump-files and text files, the full Service Database informations about probes,
boards, Software, Options and the Event Log File to the DVD Drive.
Operation see:
Section 7-4-3 "Export Log’s and System Data" on page 7-16
.
5-14-3-9
Console Test
Select the MONITOR TEST button to check control console function.
Operation see
5-14-3-10
Keyboard Layout
To change the keyboard layout to different languages.
Operation see:
Section 6-4 "Modification of Keyboard Layout" on page 6-3
.
NOTE:
Reconfigure the layout of the keyboard is only useful by changing the concerned keys also.
see:
Section 8-51 "Replacement of AN KEY TOP" on page 8-151
5-14-3-11
Monitor Test
Select the MONITOR TEST button to perform color calibration.
Operation see:
Section 6-3 "LCD Monitor Adjustment" on page 6-2
5-14-3-12
Connected Probes
Select the Connected Probes button. The “Probe Status” window supplies information about type and
serial number of connected probe and which position is used for it. There is no edit function included.
Figure 5-30 Probe Status