Chapter 10 - Care & Maintenance
10-17
Table 10-18 Type CF Patient Auxiliary Limits
Table 10-19 Isolation between Mains and Applied Limits****- Patient Leakage Main to Type BF
Applied part
Table 10-20 Isolation between Mains and Applied Limits****- Patient Leakage Main to Type CF
Applied part
Table 10-21 Total Patient leakage current: Type CF probe to ground (Only for Europe)
Conditions
Country
USA
Others
Close Protective earth and close neutral with normal polarity
0.01 mA*
0.01 mA***
Close Protective earth and close neutral with reverse polarity
0.01 mA*
0.01 mA***
Open Protective earth and close neutral with normal polarity
0.05 mA*
0.05 mA***
Open Protective earth and close neutral with reverse polarity
0.05 mA*
0.05 mA***
Open neutral and close Protective earth with normal polarity
0.05 mA*
0.05 mA***
Open neutral and close Protective earth with reverse polarity
0.05 mA*
0.05 mA***
Conditions
Country
USA
Others
Close Protective earth and close neutral with normal polarity
5 mA*
5 mA***
Close Protective earth and close neutral with reverse polarity
5 mA*
5 mA***
Conditions
Country
USA
Others
Close Protective earth and close neutral with normal polarity
0.05 mA*
0.05 mA***
Close Protective earth and close neutral with reverse polarity
0.05 mA*
0.05 mA***
Conditions
Limitation
Close Protective earth and close neutral with normal polarity
0.05 mA*****
Close Protective earth and close neutral with reverse polarity
0.05 mA*****
Open Protective earth and close neutral with normal polarity
0.1mA*****
Open Protective earth and close neutral with reverse polarity
0.1mA*****
Open neutral and close Protective earth with normal polarity
0.1mA*****
Open neutral and close Protective earth with reverse polarity
0.1mA*****
Summary of Contents for Voluson P8