Section 3-8 - System Configuration
3-8-1-10
How to activate User Login Authentication
1.) Invoke System Setup as described in
2.) On the right side of the screen select
General Settings
and then click the
Authentication
tab.
3.) Check mark
Activate User Login
.
a.) Click
Add User.
A dialog appears.
b.) Enter the
User Name
, the
First Name
and the
Last Name
.
c.) Enter a
Password
and
Confirm Password
.
NOTE:
A valid password must be at least 6 characters long and has a maximum length of 80 characters. The
password must contain at least 2 non-letter characters, 0...9 or !@#$%^().
d.) Select a
Group
the user belong to.
e.) Confirm with
OK
to save user Login Authentication.
4.) Close the Service page with
Save&Exit
and restart the system.
The system login window appears as soon as the system is started or a current user logs off. The
Emergency
button enables emergency mode with restricted access. Either ask the customer for proper
login credentials or connect the “standard GE” service dongle to omit the login.
NOTE:
For further details refer to the Voluson P6/P8 Basic User Manual.
Figure 3-57 Checking mark ‘Activate User Login’