8-134
Section 8-42 - Replacement of the DRFG
3.) Program the System Serial number and re-enter the Permanent Key
To program the System SNO, click the
OPTIONS
tab:
a.) Enter the "Serial key" (alphanumeric letters based on system serial number and DRFG
Hardware ID) in the field "Serial Number" (e.g., AEE1193428AA6078123456SU7). Then press
ENTER. it takes about several seconds to get response.
b.) Enter the "Permanent Key" in the field Permanent key, click the
SUBMIT
button and afterwards
SAVE & EXIT
.
Confirm the Message box "BT-key state has been changed." with YES. The system reboots.
4.) Check System Serial number, System Type, Hardware ID and Software options
a.) Press the
F2
key on upper OPIO, select
SYSTEM SETUP
, select
ADMINISTRATION
and then
select the
OPTION
tab.
System Type must be correct depending on Voluson™ P8/Voluson™P6.
Software options and serial number must be correct.
5.) Perform:
FRU8-29 Replacement of the DRFG - Functional Tests.
Figure 8-193 System Setup - Administration - Options page
NOTICE
!! NOTICE:
If the previous mentioned steps was done correctly, the system will show the correct SNO and the prior
installed options are enabled in the
OPTIONS
page.
NOTICE
!! NOTICE:
DO NOT
reboot the system before you have changed the serial number successfully!
Field to enter
the “Serial Key”
Field to enter
Permanent key
Summary of Contents for Voluson P8