Section 7-8 - Minimum Configuration to Boot/Scan
3.) Keep the
H
key pressed during complete startup (~ 1 minute). Following message appears:
4.) Click on
CONTINUE
. The "Probe Selection" menu appears on the screen.
5.) Press the
FREEZE
key. The application starts, but does not show any Ultrasound information.
6.) Backup all images (see:
Section 4-5-6 "Archiving Images" on page 4-21
).
7.) Save a "Full Backup" as described in
Section 4-5-3 "Save Full System Configuration (Full Backup)"
) or press the
ALT+D
key to gather detailed information and data from the system.
(see:
Section 7-4-3-1 "Export System Data (by pressing the ALT + D key)" on page 7-14
).
8.) Press the
F2
key on the control panel and select
SYSTEM SETUP
to invoke setup desktop.
9.) Select
ADMINISTRATION
, open the
SYSTEM INGO
page and click
REBOOT MACHINE
.
Figure 7-34 H Key pressed during startup
Figure 7-35 Reboot Machine