Chapter 5 - Components and Functions (Theory)
5-55
5-14-3-14
Probe EEPROM
NOTE:
It is recommended to connect and re-program one probe after the other.
1.) Click the
Probe
EEPROM
button on the “Service Tools” page (see
2.) Check and/or choose correct probe name from the pull-down menu (1).
3.) Check and/or enter probes Serial Number in the appropriate [SNO] filed(2).
4.) Select “Today” from the Start date pull-down menu(3).
5.) Click the
OK
( Global
EEPROM
) button(4).
NOTE:
DO NOT disconnect the probe during the re-programming process!
Probe ID and coding is assigned to the probe that is connected to the selected prot.
6.) Confirm with
OK
.
CAUTION
Do not assign wrong coding (e.g., programming a RAB to a RIC probe). An inaccurate probe
programming will mostly likely damage the probe and/or system.
Figure 5-30 Probe EEPROM