GE
RAFT
V
OLUSON
™ P8 / V
OLUSON
™ P6
DIRECTION 5723243, R
EVISION
6
DRAFT (M
AY
23, 2018)
B
ASIC
S
ERVICE
M
ANUAL
4-12
Section 4-5 - Backup and Restore Database, Preset Configurations and Images
4-5-1
Save Small Backup (Scan Settings)
The Image Settings contains:
•
Application Settings
•
2D Factory and 2D User Presets
•
3D/4D Factory and 3D/4D User Presets
•
Annotation Presets
•
Scan Assistant Configuration
•
Measure Configuration
•
Biopsy Lines
1.) Insert a DVD/CD+R/RW into the drive or connect an external USB device.
2.) Press the
F2
key on the control panel.
3.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen.
4.) On the left side of the screen select BACKUP and then click the
SYSTEM CONFIGURATION
tab.
5.) Click the SAVE button of the “Small Backup(Scan Settings)” group.
6.) Choose the media and click the SAVE button.
7.) Select the NEW FILE... key and enter a file name (without extension).
8.) Click the OK key to start the process. When the saving has been completed, click OK.
Figure 4-11 Image Settings Only - Save window