Section 4-7 - Peripheral Checks
Section 4-7
Peripheral Checks
Check that peripherals work as described below:
4-7-1
ECG Check Out
Connect the USB ECG and check :
Table 4-7
Peripheral Checks
Step
Task to do
Expected Result(s)
1
Press the
FREEZE
key.
Stop image acquisition.
2
Press the remote key (
P1
,
P2
,
P3
)
which is assigned to the BW printer.
The image displayed on the screen is printed on the Black & White
printer.
3
Press the remote key (
P1
,
P2
,
P3
)
which is assigned to the color printer.
The image displayed on the screen is printed on the Color printer.
4
Press the remote key (
P1
,
P2
,
P3
)
which is assigned to recorder control.
Recording starts/stops.
Step
Task
Expected Result(s)
1
Press the F2 key on the control panel and then
press the “ECG Line” button to display the ECG
signal.
It will display a curve along the bottom edge of the image sector.