Chapter 3 - Setup Instructions
3-67
3-9-1-6
How to change Ext. Monitor Output Settings
1.) Invoke system Setup as described in
2.) On the right side ofr the screen select Connectivity and then click the Peripherals tab.
3.) If not currently selected, click the proper field.
4.) Close the Service page with Save&Exit and restart the system.
3-9-1-7
How to adjust function of the Footswitch (Left/Middle/Right)
1.)
Invoke System Setup as described in Section 3-8-1-1 on page 3-55
.
2.) On the left side of the screen select CONNECTIVITY and then click on the
PERIPHERALS tab.
3.) Select desired function of the Footswitch Left, Middle and Right.
4.) Click SAVE&EXIT to save Settings and exit System Setup.
3-9-1-8
How to change the Keyboard Layout
Chapter 6 - Modification of Keyboard Layout, on page 6-3
.
3-9-1-9
How to configure Service Platform
Chapter 3 - How to configure Service Platform, on page 3-67
3-9-1-10
How to install SW Option
For instruction, refer to Voluson S10/S8/S6 and P6/P8 Software Option installation manual, 5724006 in
CDL and CDS.
Figure 3-73 System setup -Connectivity - Peripherals page
Footswitch