Section 8-6 - Mechanical Platform Components - Replacement Procedures
P R E L I M I N A R Y
8-6-9
Gas Spring Replacement Procedure
8-6-9-1
Tools
Use the appropriate Phillips screwdriver, light hammer, punch, and 5 mm Allen key.
8-6-9-2
Time Required
120 min
8-6-9-3
Preparation
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
.
8-6-9-4
Gas Spring Removal Procedure
1) Perform the following removal procedures:
2) Unscrew the shaft covering screws located on either side of the main column - see
FRU Part #
.
•
21.5" Monitor Replacement Procedure on page 8 - 31
•
Operator Panel Keyboard Assembly Removal Procedure on page 8 - 44
•
OPIO Basket Removal Procedure on page 8 - 26
(if attached to the system)
Figure 8-194 Shaft Covering Screws
Summary of Contents for Vivid S60N
Page 4: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ii ...
Page 5: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iii ...
Page 6: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iv ...
Page 7: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL v ...
Page 8: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vi ...
Page 9: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vii ...
Page 10: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL viii ...
Page 11: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ix ...
Page 12: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL x ...
Page 13: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xi ...
Page 14: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xii ...
Page 477: ......