Chapter 3 - System Setup
3-51
P R E L I M I N A R Y
Section 3-9
Options Setup
3-9-1
Software Options
Most of the options for Vivid™ S60/Vivid™ S70 are activated by installing a password (alphanumeric
text string).
•
For installation instructions, see:
Software Options Configuration on page 3 - 39
3-9-2
Color Video Printer Setup
NOTE:
The Color Video Printer is an option.
An external Color Video Printer may be connected to one of the USB ports on Vivid™ S60/Vivid™ S70.
NOTE:
SONY UPD-2X-MD is a medical device so it can be placed near the Vivid™ S60/Vivid™ S70. The USB
cable is 5 meters long, so the printer must be placed within reach of this cable.
Follow these steps to connect the printer to the Vivid™ S60/Vivid™ S70:
1) Connect the USB cable to the printer and to one of the USB ports on Vivid™ S60/Vivid™ S70.
2) Select the correct Power cable and connect it to the printer and to the mains power outlet.
3) Switch on the power switch on the printer.
NOTE:
To be able to use the printer, the printer must be selected on the Vivid™ S60/Vivid™ S70.
3-9-3
USB Flash Card Setup
There is no special setup procedure for use of a USB Flash Card.
3-9-4
Wireless Network Adaptor
The Wireless Network adaptor should be connected to the lower USB socket on the Vivid S60N/Vivid
S70N Rear I/O Peripheral/Accessory Connector Panel.
NOTE:
It is possible to connect the Wireless Network adaptor either once the system is powered ON, or after
shutdown.
All software drivers for the Wireless Network are pre-installed for the designated Wireless Network
adaptor only.
WARNING
WARNING
WHEN CONNECTING PERIPHERALS TO THE VIVID™ S60/VIVID™ S70 SYSTEM,
UNLESS THE PERIPHERAL IS MEDICAL-GRADE EQUIPMENT IT SHOULD BE
CONNECTED TO THE ISOLATED USB CONNECTOR.
NOTICE
IMPORTANT
During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
WARNING
WARNING
DO NOT ATTEMPT TO USE A DIFFERENT TYPE OF WIRELESS NETWORK ADAPTOR.
THE ULTRASOUND SYSTEM IS AN EXTREMELY SENSITIVE AND COMPLEX
MEDICAL SYSTEM. ANY UNAUTHORIZED PERIPHERALS MAY CAUSE SYSTEM
FAILURE OR DAMAGE!
Summary of Contents for Vivid S60N
Page 4: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ii ...
Page 5: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iii ...
Page 6: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iv ...
Page 7: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL v ...
Page 8: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vi ...
Page 9: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vii ...
Page 10: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL viii ...
Page 11: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ix ...
Page 12: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL x ...
Page 13: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xi ...
Page 14: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xii ...
Page 477: ......