Chapter 10 - Care and Maintenance
10-27
P R E L I M I N A R Y
Follow these steps to test each probe for leakage current:
1) Turn the Vivid™ S60/Vivid™ S70 OFF.
2) Plug the Ultrasound system’s mains power cord into the test meter, and plug the test meter into the
tested AC wall outlet.
3) Plug the Chassis Ground Probe (saline probe) into the test meter's “CHASSIS” connector.
4) Using any calibrated Electrical Safety Analyzer tool compliant with AAMI/ESI 1993 or IEC 60601 or
AS/NZS 3551, switch the meter's function selector to the LEAD-LEAD position (or equivalent
function in the device).
5) Connect the ultrasound probe to be tested to the Vivid™ S60/Vivid™ S70.
6) Put the saline probe and the ultrasound probe’s probe face (imaging area of the probe) into the
saline bath.
7) Power ON the Vivid™ S60/Vivid™ S70.
8) After the Vivid™ S60/Vivid™ S70 has completed the boot process, select the probe to be tested so
it is the active probe.
9) Depress the LIFT GROUND rocker switch and record the highest current reading.
10) Follow the test conditions and test limits described in:
•
or
•
as applicable for every probe.
The test passes when all readings measure less than the stated limits.
11) Keep a record of the results with other hard copies of maintenance data using
page 10-28.
Table 10-17
shows a typical format for recording ultrasound probe source leakage current.
Measurements should be recorded for each probe under the set of test conditions specified in:
•
or
•
as applicable.
NOTE:
Values in italics font are given as examples only.
12) Record all data on the Inspection Certificate.
DANGER
TO AVOID PROBE DAMAGE AND POSSIBLE ELECTRIC SHOCK, DO NOT IMMERSE
PROBES INTO ANY LIQUID BEYOND THE LEVEL INDICATED IN THE PROBE USERS
MANUAL.
DO NOT TOUCH THE PROBE, CONDUCTIVE LIQUID OR ANY PART OF THE
UNIT UNDER TEST WHILE DOING THE TEST
.
CAUTION
Equipment damage possibility. Never switch the Polarity or the status of the Neutral
when the Ultrasound system is powered on. Power off the Ultrasound system, allow the
stored energy to bleed down, and turn the circuit breaker off BEFORE switching the
POLARITY switch and/or the NEUTRAL switch on the leakage meter to avoid possible
power supply damage.
Summary of Contents for Vivid S60N
Page 4: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ii ...
Page 5: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iii ...
Page 6: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iv ...
Page 7: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL v ...
Page 8: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vi ...
Page 9: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vii ...
Page 10: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL viii ...
Page 11: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ix ...
Page 12: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL x ...
Page 13: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xi ...
Page 14: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xii ...
Page 477: ......