Chapter 8 - Replacement Procedures
8-173
P R E L I M I N A R Y
Section 8-7
Operating System and/or Application Software Loading Procedures
8-7-1
Software Installation Procedures - General Overview
The Vivid™ S60/Vivid™ S70 ultrasound scanner software loading options provide the ability to install
the Windows operating system, the Vivid™ S60/Vivid™ S70 application software - or both - depending
on current requirements.
The complete installation procedure can be performed using the Touch Screen only. An on-screen
(virtual) keyboard is available whenever text input is required. In addition, the optional A/N Keyboard
can be used - or any regular USB keyboard.
When installing the Windows operating system software, in some situations (for example, after
performing a
Hard Disk Replacement Procedure
), it is necessary to format the hard disk.
Be aware that this will perform a full format of the Hard Disk and
all patient data will be lost
.
Normally, the Vivid™ S60/Vivid™ S70 system is supplied with the software already installed. In some
cases, it is necessary to re-install the software.
NOTE:
After performing a software installation, all logs are deleted. It is recommended that these
should be recorded should they be needed in the future.
If you are upgrading the system software from a previous version, it is necessary to contact your local
OTR department to receive an appropriate software password. Make sure you specify the system serial
number located on the Vivid™ S60/Vivid™ S70 unit.
WARNING
WARNING
DO NOT ATTEMPT TO INSTALL SOFTWARE THAT WAS NOT DESIGNATED FOR YOUR
VIVID™ S60/VIVID™ S70 UNIT. ATTEMPTING TO INSTALL UN-APPROVED SOFTWARE
WILL CAUSE IRREVERSIBLE DAMAGE TO HARDWARE AND SOFTWARE!
NOTICE
IMPORTANT Since neither the System Serial Number or the Computer Name can be changed after the
software installation procedure is completed, it is important to make sure that when prompted, the
correct information is inserted during the installation procedure.
WARNING
WARNING
WHILE THE SOFTWARE INSTALLATION PROCEDURE IS DESIGNED TO PRESERVE
DATA, YOU SHOULD SAVE ANY PATIENT DATA, IMAGES, SYSTEM SETUPS TO
BACKUP MEDIA BEFORE DOING A SOFTWARE INSTALLATION.
NOTICE
IMPORTANT Before performing any Vivid™ S60/Vivid™ S70 software installation procedure, it is
mandatory to backup the Archive. Follow the instructions as described in
Installation Procedures on page 8 - 174
.
Summary of Contents for Vivid S60N
Page 4: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ii ...
Page 5: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iii ...
Page 6: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iv ...
Page 7: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL v ...
Page 8: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vi ...
Page 9: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vii ...
Page 10: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL viii ...
Page 11: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ix ...
Page 12: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL x ...
Page 13: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xi ...
Page 14: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xii ...
Page 477: ......