D
IRECTION
FR091521, R
EVISION
1
V
IVID
S60N/V
IVID
S70N
B
ASIC
S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-153
P R E L I M I N A R Y
8-6-6
LCD Arm Riser for 21.5 inch monitor replacement procedure
8-6-6-1
Tools
Use the appropriate Phillips type screwdrivers, side cutting pliers and a 5mm Allen key.
8-6-6-2
Time required
90 minutes
8-6-6-3
Preparations
1.) Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
4 - 7
.
2.) Remove the
21.5in Monitor
Remove the 21.5 inch Monitor on page 8 - 31
3.) Unscrew and remove the
three fastening screws
. See the illustration below.
4.) Remove the
two covers
from the LCD Arm Riser. See the illustration below.
5.) Remove the
LCD Arm Alton 21.5in Wide
as described in
LCD Arm Alton 21.5in Wide Monitor
Removal Procedure on page 8 - 150
.
Figure 8-186 LCD Arm Riser for 21.5 inch monitor
FRU Part #
on page 9-3. Look for: LCD Arm Riser for 21.5 inch monitor
Figure 8-187 Remove the covers from the LCD Arm Riser
Summary of Contents for Vivid S60N
Page 4: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ii ...
Page 5: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iii ...
Page 6: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iv ...
Page 7: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL v ...
Page 8: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vi ...
Page 9: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vii ...
Page 10: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL viii ...
Page 11: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ix ...
Page 12: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL x ...
Page 13: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xi ...
Page 14: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xii ...
Page 477: ......