Section 8-6 - Lower Section Components Replacement
8-6-7
Up/Down/Swivel Handle Replacement Procedure
NOTE:
This section describes the removal and installation procedures for the metal up/down/swivel handle
(Part No. 237002).
8-6-7-1
Tools
Use the appropriate the Phillips-type screw drivers, Allen key, and combination wrench, as indicated in
the up/down/swivel handle removal procedure.
8-6-7-2
Preparation
Shut down the Vivid™ 4
ultrasound unit, as described in
.
8-6-7-3
Up/Down/Swivel Handle Removal Procedure
1) Remove the control console lower cover rear, as described in the
2) Remove the control console lower cover, as described in the
3) Remove the control console upper cover front section, as described in the
Cover (Front) Removal Procedure
4) Disconnect the four
keyboard matrix cables as shown in
Figure 8-82 Vivid™ 4 Up/Down/Swivel Handle in Position
Figure 8-83 Disconnecting the Keyboard Matrix Cables
Up/Down/Swivel Handle
(used for control console
up/down/swivel
movement)