GE M
EDICAL
S
YSTEMS
D
IRECTION
2392751-100, R
EVISION
3
V
IVID
™ 4 S
ERVICE
M
ANUAL
7-76
Section 7-4 - Performing Back End Diagnostics on the System
7-4-4-3
Manual Lights Test
1) Access the
HW Diagnostics
menu, as described in the
Accessing the Back End Diagnostic Options
2) Trackball to the
External Keyboard
button and press
Select
. The
Extended Keyboard Test
window
is displayed, as shown in
3) In the
Test Manager
tab, trackball to the
Lights (Manual)
radio button and press
Select
.
4) In the
KB Panel
tab, deselect the
Select
checkbox. Verify that after the
Select
LED is off, the
alphanumeric keyboards LEDs are dimmed.
5) When you are finished, trackball to the
Exit
button and press
Select
.
7-4-4-4
Commands History
Select the
Commands History
tab in the
Extended Keyboard Test
window to view a report of the
activity of the extended control panel during the test, as shown below:
Figure 7-50 Commands History Tab - Extended Keyboard Test Window
7-4-4-5
General Configuration
Select the
General Configuration
tab in the
Extended Keyboard Test
window to set the type of log files
that is generated, for example an error log or no log at all.