GE M
EDICAL
S
YSTEMS
D
IRECTION
2392751-100, R
EVISION
3
V
IVID
™ 4 S
ERVICE
M
ANUAL
8-52
Section 8-5 - Back End Parts Replacement
21.) If available on the system, connect the Network cable (Part No. 2292972) to the Network socket [
9
].
22.) Connect the Back End Audio BKT EXT cable (Part No. 2258803) to the Audio IN/OUT socket [
10b
].
23.) Connect the Back End Audio BKT EXT cable (Part No. 2258804) to the Audio IN/OUT socket [
10a
].
24.) Connect the PC2IP cable (Part No. 2276874) to the PC2IP IN/OUT socket [
11
].
25.) If available on the system, connect the modem communication cable (Part No. 2320338) to the
COM 3 and 4 socket [
12
],
OR
connect the frame grabber cable (Part No. 2371429) to the socket [
12
].
26.) Connect the Back End KB Cont P1 cable (Part No. 2252975) to the Ext Keyboard socket [
13b
].
27.) Connect the Back End KB Cont P2 cable (Part No. 2252976) to the Ext Keyboard socket [
13a
].
28.) Connect the VIC/ Rear Panel left connector cable (Part No. 2253011) to the RGB Video OUT
socket [
14
].
29.) Connect the Keyboard/VIC External power cable (Part No. 2252976) to the power to Keyboard
Matrix socket [
15
].
Note: IMPORTANT -
Do NOT connect the HASP key to this socket!
- refer to
above
[connection to socket
7
].
30.)
For systems
without
a UPS
:
Connect the BEP power cable (Part No. 2276876) to the power input socket [
16a
]
For systems using a UPS
:
Connect the UPS power cable (Part No.2307289) to the Plug and Scan socket [
16b
]
and then connect the BEP/Modem power cable (Part No. 2300847) to the power input socket [
16a
].
31) Facing the side cover, push the BEP as far to the right as possible, and tighten first the left lower
bracket, then the right lower bracket and finally the bracket on its upper left side. Use two screws
to fasten each bracket.
32) Turn ON the system and boot-up in
Technical Support
mode - make sure the T-plug (dongle) is
inserted into the parallel or USB port on the Vivid™ 4 rear connectors panel.
33.)Re-install the Software - see the
Software Installation/Upgrade Procedure
The system will perform four reboots before completing the software installation, after which the
probe select screen will be displayed.
Note:
When replacing a BEP, one must re-install software using the software provided with the
system.
Do NOT use the software supplied with the BEP unless it is a version later
than the software currently installed on the system
.
Note:
IMPORTANT: During the software installation, the system will display a few messages;
reply to all by selecting
OK
, or
Cancel
(all can be ignored).
34.)After a complete installation of the software (in accordance with
), boot-up the system.
35.)Press the
Config
button on the control console; the System Configuration dialog box is displayed.
36.)Trackball to the
Technical Support
tab and press
Select
.
Table 8-3 Software Installation Compatibility
Vivid 4
RFT
Vivid 4
RFI
SW Version 4.3
SW Version 4.6