GE M
EDICAL
S
YSTEMS
D
IRECTION
2392751-100, R
EVISION
3
V
IVID
™ 4 S
ERVICE
M
ANUAL
7-116
Section 7-5 - Common Service Interface
•
Clean UserDefs
-
used to clean all user-defined settings and restore the default
settings
Figure 7-90 Clear User-defined Settings
CAUTION:
THIS PROCESS IS IRREVERSIBLE! Before proceeding, make sure you have a
complete backup of the current settings if they will be required in the future
.
When selecting
Clean Userdefs
, the following warning message is displayed:
After making a back-up copy of the current settings, click
OK
to proceed.
NOTE:
This function performs the same procedure as that of Restore Factory Presets (accessed under the
System Configuration window - System tab).