background image

GE M

EDICAL

 S

YSTEMS

 

D

IRECTION

 2392751-100, R

EVISION

 

3

V

IVID

™ 4 S

ERVICE

 M

ANUAL

8-76

Section 8-5 - Back End Parts Replacement

8.) From the rear of the BEP2, remove the four screws that fasten the power supply in position on the 

BEP, as shown in 

Figure 8-59

.

Note:

For the next steps, it is recommended to carefully lie the BEP2 on its side, to provide easier 

access.

9.) Cut any cable ties holding the power supply cable in place, as shown in 

Figure 8-60

 below.

Figure 8-59 BEP2 Rear View

Figure 8-60 BEP2 Power Supply and Cable Ties

Power Supply

Cut Cable Ties Here

Cut Cable Ties Here

Summary of Contents for Vivid 4 BT03 RFI

Page 1: ...Technical Publication Part Number 2392751 100 Revision 3 GE Medical Systems Vivid 4 Service Manual Copyright 2006 by GE Medical Systems...

Page 2: ...UE L ANGLAIS C EST AU CLIENT QU IL INCOMBE DE LE FAIRE TRADUIRE NE PAS TENTER D INTERVENTION SUR LES QUIPEMENTS TANT QUE LE MANUEL SERVICE N A PAS T CONSULT ET COMPRIS LE NON RESPECT DE CET AVERTISSEM...

Page 3: ...NUAIS NOUTRO IDIOMA DA RESPONSABILIDADE DO CLIENTE FORNECER OS SERVI OS DE TRADU O N O TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E COMPREENDIDO ESTE MANUAL DE ASSIST NCIA T CNICA O N O CUMPRIMENT...

Page 4: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL iii...

Page 5: ...tors Other connections between pieces of electrical equipment calibrations and testing shall be performed by qualified GE Medical Systems personnel In performing all electrical work on these products...

Page 6: ...ted in any form in whole or in part without prior written permission of GE Medical Systems GE Medical Systems may revise this publication from time to time without written notice TRADEMARKS All produc...

Page 7: ...ge N A Chapter 2 Pre Installation pages 2 1 to 2 12 3 Chapter 7 Diagnostics Troubleshooting pages 7 1 to 7 132 3 Important Precautions pages i to iv 3 Chapter 3 Installation pages 3 1 to 3 80 3 Chapte...

Page 8: ...s 1 4 Important Conventions 1 5 Conventions Used in this Manual 1 5 Model Designations 1 5 Icons 1 5 Safety Precaution Messages 1 5 Standard Hazard Icons 1 6 Safety Considerations 1 7 Introduction 1 7...

Page 9: ...IRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL viii Electrostatic Discharge ESD Prevention 1 24 Standards Used 1 25 Lockout Tagout Requirements For USA Only 1 25 Customer Assistance 1 26 Conta...

Page 10: ...ite Circuit Breaker 2 3 Site Power Outlets 2 4 Mains Power Plug 2 4 Power Stability Requirements 2 4 EMI Limitations 2 5 Probe Environmental Requirements 2 6 Facility Needs 2 7 Purchaser Responsibilit...

Page 11: ...paring for Installation 3 13 Confirming Customer Order 3 13 Verifying the Shipping Crate Contents 3 13 External Cables 3 13 Accessories 3 14 Component Inspection 3 14 Damage Inspection Checklist 3 14...

Page 12: ...s Tab 3 40 System Options Tab 3 42 Printers Tab 3 44 VCR ECG Tab 3 45 Technical Support Tab 3 47 Technical Support Tab RFI Configuration 3 47 Technical Support Tab RFT Configuration 3 49 Technical Sup...

Page 13: ...pping Crate and Packaging Materials 3 71 Cardboard Shipping Carton and Packaging Materials 3 72 Packing the Unit into the Wooden Shipping Crate 3 73 Assembling the Wooden Shipping Crate 3 74 Packing t...

Page 14: ...rinter Tests 4 5 Mechanical Functions 4 6 Console Movement Test 4 6 Brakes Function Test 4 6 Back End Processor Tests 4 7 Internal ECG Test 4 7 Phono Test optional 4 7 Magneto Optical Drive MOD Test 4...

Page 15: ...739L Probe Image Quality Tests 4 21 2D Noise Uniformity Test 4 21 2D Performance Test 4 21 CFM Stationary Noise Test 4 22 Probe Operation Check Alignment Sensitivity 4 22 CFM Operation Tests 4 23 PW O...

Page 16: ...I configuration only 5 12 Transmit Signal Path 5 12 Received Signal Path 5 12 Front Board Assembly FB 5 16 General Description 5 16 TR4 Board Description 5 17 MUX Board 5 18 General Description 5 18 B...

Page 17: ...5 38 Floppy Drive 5 39 Hard Disk 5 39 Magneto Optical Drive MOD 5 40 CD Read Write CDRW 5 40 ECG Module 5 40 Modem 5 41 PC VIC Assembly 5 42 General Description 5 42 PAL 5 43 NTSC 5 43 External Periph...

Page 18: ...vice Platform 5 52 Introduction 5 52 iLinq Interactive Platform Features 5 52 Web Server Browser 5 52 Connectivity 5 52 Configuration 5 52 Contact GE 5 52 Interactive Application 5 52 Global Service U...

Page 19: ...rt IMP LEDs 6 5 Front End Controller FEC LEDs 6 6 RF and Tissue Processor RFT 6 6 Beamformer BF 6 6 Channels Multiplexer MUX 6 7 Front Board Assembly FB 6 7 Back End Power Supply Voltages 6 8 VIC Vide...

Page 20: ...c Tests for RFT Configuration 7 15 RFT Diagnostic Tests for RFT Configuration 7 17 Front End Controller FEC Diagnostic Tests for RFT Configuration 7 19 Beamformer BF Diagnostic Tests 7 21 Front Board...

Page 21: ...ry 7 76 General Configuration 7 76 Keyboard Diagnostic Test 7 77 Media Driver Diagnostic Test 7 78 MOD Diagnostic Test 7 78 CD Writer Diagnostic Test 7 79 Computer Diagnostic Test 7 80 Benchmark Testi...

Page 22: ...7 106 Monitor Test 7 106 Front End Diagnostics 7 107 Image Quality Page 7 111 Calibration Page 7 111 Configuration Page 7 112 Utilities Page 7 113 Windows 2000 Utility 7 113 Scanner Utilities 7 115 P...

Page 23: ...re 8 6 Rear Cover Replacement Procedures 8 7 Tools 8 7 Preparations 8 7 Rear Cover Removal Procedure 8 7 Rear Cover Installation Procedure 8 7 Connector Panels Cover Replacement Procedures 8 8 Tools 8...

Page 24: ...r Removal Procedure 8 19 Control Console Lower Rear Cover Installation Procedure 8 19 Front Handle Replacement Procedure 8 20 Tools 8 20 Preparation 8 20 Front Handle Removal Procedure 8 20 Front Hand...

Page 25: ...eaker Replacement Procedure 8 34 Tools 8 34 Preparation 8 34 Speaker Removal Procedure 8 34 Speaker Installation Procedure 8 34 Front End Parts Replacement 8 35 Front End Boards Replacement Procedure...

Page 26: ...ure 8 54 BEP2 Side Cover Installation Procedure 8 54 BEP2 Front Cover Removal Procedure 8 54 BEP2 Front Cover Installation Procedure 8 56 BEP2 Octopus Card Holder Removal Procedure 8 56 BEP2 Octopus C...

Page 27: ...ols 8 66 Preparation 8 66 CDRW Drive Removal Procedure 8 66 CDRW Drive Installation Procedure 8 67 MO Drive Replacement Procedure 8 68 Tools 8 68 Preparation 8 68 MO Drive Removal Procedure 8 68 MO Dr...

Page 28: ...cedure 8 88 AC BEP or FE Cable Installation Procedure 8 88 Gas Spring Cable Replacement Procedure 8 89 Preparation 8 89 Gas Spring Cable Removal Procedure 8 89 Gas Spring Cable Installation Procedure...

Page 29: ...parations 8 118 Mitsubishi VCR Installation Procedure 8 119 Mitsubishi VCR Setup Procedure 8 122 Mitsubishi VCR Removal Procedure 8 125 Sony VCR Replacement Procedure 8 126 Tools 8 126 Manpower 8 126...

Page 30: ...MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL xxix HP 6540 3 USB Deskjet Color Printer Installation Procedure 8 140 HP 6540 3 USB Deskjet Color Printer Removal Procedure 8 1...

Page 31: ...Renewal Parts Overview 9 1 Purpose of Chapter 9 9 1 List of Abbreviations 9 2 Renewal Parts Lists and Diagrams 9 3 Mechanical Hardware Parts 9 3 AC System Parts 9 9 Front End Parts 9 11 Back End Part...

Page 32: ...ance 10 6 System Periodic Maintenance 10 7 Preliminary Checks 10 7 Functional Checks 10 8 System Checks 10 8 Peripheral Option Checks 10 9 Input Power Checks 10 9 Mains Cable Inspection 10 9 Cleaning...

Page 33: ...Lead Source Leakage Lead to Ground 10 24 Definition 10 24 Generic Procedure 10 24 Isolated Patient Lead Source Leakage Lead to Lead 10 25 Isolated Patient Lead Sink Leakage Isolation Test 10 25 Data...

Page 34: ...Chapter 3 Installation Contains installation procedures and an installation checklist Chapter 4 Functional Checks Contains functional checks that are recommended as part of the installation procedure...

Page 35: ...Parts Contains a complete list of field replaceable parts for the Vivid 4 ultrasound unit Chapter 10 Periodic Maintenance Provides periodic maintenance procedures for the Vivid 4 ultrasound unit 1 1 3...

Page 36: ...and below are configured with RFT hardware Table 1 2 Vivid 4 BT03 RFI Models Model Cat No Description BASE Vivid 4 Console Vivid 4 BT03 console 220 240V AC RFI H45011FT An advanced version of the Ultr...

Page 37: ...und unit and therefore shares much of the same hardware However even though much of the hardware is the same the Vivid 3 cannot be upgraded to Vivid 4 or vice versa The Vivid 4 ultrasound unit provide...

Page 38: ...ls of safety precautions are found on the equipment and throughout this service manual Different levels of severity are identified by one of the following icons which precede precautionary statements...

Page 39: ...eral different graphical icons symbols may be used to make you aware of specific types of hazards that could cause harm Other hazard icons make you aware of specific procedures that should be followed...

Page 40: ...HAT CAN EASILY BE DAMAGED BY IMPROPER HANDLING USE CARE WHEN HANDLING AND PROTECT FROM DAMAGE WHEN NOT IN USE DO NOT USE A DAMAGED OR DEFECTIVE PROBE FAILURE TO FOLLOW THESE PRECAUTIONS CAN RESULT IN...

Page 41: ...e shared with other types of equipment Both the system power cable and the power connector must meet international electrical standards Note When replacing UPS batteries ensure that defective batterie...

Page 42: ...DANGEROUS VOLTAGES CAPABLE OF CAUSING DEATH ARE PRESENT IN THIS EQUIPMENT USE EXTREME CAUTION WHEN HANDLING TESTING AND ADJUSTING EXPLOSION WARNING DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSP...

Page 43: ...rtinent information regarding the operation of the ultrasound unit 1 4 1 Product Label Locations The following two diagrams indicate the location of some of the labels and icons found on the Vivid 4 u...

Page 44: ...EDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 1 Introduction 1 11 Figure 1 2 Product Label and Icon Locations Rear 1 Main Label 2 AC Voltage Rating Label 3 GND Label 1...

Page 45: ...cannot become live in the event of a failure of the basic insulation Rear of the unit and probe connectors Device Listing Certification Labels Laboratory logos or labels that denote conformity with in...

Page 46: ...the lightning flash with arrowhead in equilateral triangle is used to indicate electric shock hazards Rear of the unit Protective Earth Indicates the protective earth grounding terminal Rear of the un...

Page 47: ...Label section on page 1 23 Swivel Brake Label section on page 1 23 1 4 3 1 Main Label The main label may be printed in any of the following languages English German French Spanish Portuguese Italian C...

Page 48: ...ION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 1 Introduction 1 15 German Used in all German language countries Figure 1 5 Main Label German French Used in all French language countries Fig...

Page 49: ...REVISION 3 VIVID 4 SERVICE MANUAL 1 16 Section 1 4 Product Labels and Icons Spanish Used in all Spanish language countries Figure 1 7 Main Label Spanish Portuguese Used in all Portuguese language cou...

Page 50: ...SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 1 Introduction 1 17 Figure 1 9 Main Label Portuguese 220 240V Italian Used in all Italian language countries Figure 1 10 Main La...

Page 51: ...1 100 REVISION 3 VIVID 4 SERVICE MANUAL 1 18 Section 1 4 Product Labels and Icons Chinese Used in all Chinese language countries Figure 1 11 Main Label Chinese Danish Used in all Danish language count...

Page 52: ...ON 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 1 Introduction 1 19 Dutch Used in all Dutch language countries Figure 1 13 Main Label Dutch Finnish Used in all Finnish language countries Figu...

Page 53: ...1 100 REVISION 3 VIVID 4 SERVICE MANUAL 1 20 Section 1 4 Product Labels and Icons Greek Used in all Greek language countries Figure 1 15 Main Label Greek Japanese Used in all Japanese language countri...

Page 54: ...2751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 1 Introduction 1 21 Norwegian Used in all Norwegian language countries Figure 1 17 Main Label Norwegian Russian Used in all Russian language countrie...

Page 55: ...d in all Swedish language countries Figure 1 19 Main Label Swedish 1 4 3 2 Rating Labels Indicates the ultrasound unit s factory preset input AC voltage as follows AC 100V AC 120V AC 220 240V AC 220 2...

Page 56: ...es the locked pedal position which locks the front castors and prevents the ultrasound unit from moving The parking label shown below is located on the brake pedal at the front of the unit as shown in...

Page 57: ...produces unacceptable levels of emission This interference is often referred to as radio frequency or electromagnetic interference RFI EMI and can be radiated through space or conducted over interconn...

Page 58: ...l is required in order to achieve full EMC performance 1 5 4 Lockout Tagout Requirements For USA Only Follow OSHA Lockout Tagout requirements by ensuring you are in total control of the electrical Mai...

Page 59: ...ltrasound Service Engineering 4855 W Electric Avenue Milwaukee WI 53219 Customer Answer Center Phone 1 800 437 1171 Phone 1 800 321 7937 Phone 1 800 682 5327 Phone 1 262 524 5698 Fax 1 414 647 4125 La...

Page 60: ...he installation of a Vivid 4 ultrasound unit Included are descriptions of the electrical and facility requirements that must be met by the purchaser A worksheet is provided at the end of this chapter...

Page 61: ...if the room lighting is subdued and indirect when a scan is being performed Therefore a combination lighting system dim bright is recommended Keep in mind that lighting controls and dimmers can be a...

Page 62: ...ease note that image artifacts can occur if at any time within the facility the Ground from the main facility s incoming power source to the Ultrasound unit is only a conduit 2 2 5 1 Vivid 4 Power Req...

Page 63: ...t all current local state and national electrical codes 2 2 5 5 Mains Power Plug If the unit arrives without a power plug or with the wrong plug contact your GE dealer When necessary the installation...

Page 64: ...wever there is no guarantee that interference will not occur in a particular installation Note Possible EMI sources should be identified before the unit is installed and should not be on the same line...

Page 65: ...oose or missing covers or RF gaskets allow radio frequencies to interfere with the ultrasound signals Replace broken RF gaskets If more than 20 or a pair of the fingers on an RF gasket are broken repl...

Page 66: ...alibration and testing must also be performed by qualified personnel The products involved and the accompanying electrical installations are highly sophisticated and special engineering competence is...

Page 67: ...the case of a network option An active network outlet in the vicinity of the ultrasound unit A network cable of appropriate length regular Pin to Pin network cable An IT administrator who will assist...

Page 68: ...rator with clinically useful features for moving images and patient information over a hospital network Examples of DICOM services include the transfer of images to workstations for viewing or transfe...

Page 69: ...ork administrator must provide the required information which should include the following Vivid 4 Details DICOM network details for the Vivid 4 unit including the host name local port IP address AE t...

Page 70: ...Destination Devices TCP IP Settings Device Type Manufacturer Name IP Address Port AE Title 1 2 3 4 5 6 7 8 9 10 11 12 Dept Vivid 4 SN Type Floor Room REV Comments CONTACT INFORMATION Name Title Phone...

Page 71: ...from the basic installation Make sure system and probes have been subject to acclimation period Environmental cooling is sufficient Lighting is adjustable to adapt to varying operational conditions of...

Page 72: ...epared as described in Chapter 2 Pre Installation Once the site has been prepared installation can proceed as described in this chapter Table 3 1 Contents in Chapter 3 Section Description Page Number...

Page 73: ...troubleshooting 3 2 2 1 System Acclimation Time Following transport the Vivid 4 system may be very cold or hot Allow time for the system to acclimate before being switched ON Acclimation requires 1 ho...

Page 74: ...connect the unit to a properly grounded power outlet Do NOT use a three prong to two prong adapter as this defeats safety grounding CAUTION Do NOT wear the ESD wrist strap when you work on live circui...

Page 75: ...ge 3 9 3 3 1 Unpacking the Wooden Shipping Crate The Vivid 4 ultrasound unit is packed in a wooden crate that has four walls left right front and rear the crate base and the top cover Each section has...

Page 76: ...either the Drop or Tilt indicators have turned red showing failure please inform the GE Medical Systems sales representative immediately In addition mark on the shipping consignment note or packing sl...

Page 77: ...the probes and peripheral options items D and E in Figure 3 3 below from the wooden crate 5 Remove the console control support back foam item 11 in Figure 3 3 above 6 Remove the keyboard support foam...

Page 78: ...ntrol console monitor by pressing the release grip located under the unit s front handle 15 Remove the two cartons containing the external cables and accessories located under the control console and...

Page 79: ...wheels from swiveling as described in the Swivel Brake Label section page 1 23 18 Pull the unit backwards from the platform base onto the ramp and roll it down to the floor 19 Press the brake pedal lo...

Page 80: ...on the Shock watch and Tilt watch labels respectively for evidence of accidental shock or tilting of the crate during transit refer to Figure 3 2 on page 3 5 NOTE If the crate is damaged or if either...

Page 81: ...ews item 16 so that the cardboard carton item 12 is free of the wooden platform item 4 and the narrow flap is free 4 Open the narrow side flap of the carton remove the ramp item 5 and attach it to the...

Page 82: ...SERVICE MANUAL Chapter 3 Installation 3 11 8 Pull out and remove the monitor support item 10 in Figure 3 9 below from under the base of the monitor Figure 3 8 Front and Side View of the Unit in the Sh...

Page 83: ...se cartons contain the external cables and accessories 12 Lower the control console to its lowest position 13 Remove the manuals item 4 in Figure 3 8 on page 3 11 from the compartment on the left side...

Page 84: ...tallation 3 4 2 1 External Cables The ECG cable item number 10 or 11 is supplied according to installation location Table 3 4 Vivid 4 External Cables Item Number P N Description Quantity 1 2269430 PWR...

Page 85: ...erial number printed on the main label into the system installation details card as described in System Installation Details on page 3 78 3 4 3 1 Damage Inspection Checklist Visually inspect the conte...

Page 86: ...and peripheral fans are operating and clean 7 BE Rear Panel Check the BE rear panel connectors for bent pins loose connections and loose or missing hardware Screw all the cable connectors tightly to t...

Page 87: ...de of the front panel 4 Control Panel Contains the alphanumeric keyboard and the buttons used to operate the ultrasound unit 5 Front Handle 6 Raise Lower the Control Console Up Down Swivel Handle Loca...

Page 88: ...I O module that enables keyboard commands to be operated by foot 11 Gel Holders Situated on either side of the control console and on the top surface on either side of the monitor 12 Color Printer or...

Page 89: ...Unit The following figure shows the Vivid 4 ultrasound unit components that are visible from the rear of the ultrasound unit 1 Monitor 2 Gel Wells 3 Monitor Connection Panel 4 Left Rear Panel 5 Right...

Page 90: ...stem WARNING CONNECTING A Vivid 4 SCANNER TO THE WRONG VOLTAGE LEVEL WILL MOST LIKELY DESTROY THE SCANNER 3 4 4 2 Confirming System Voltage Configuration 1 Turn ON the system 2 In regular 2D Scanning...

Page 91: ...100 REVISION 3 VIVID 4 SERVICE MANUAL 3 20 Section 3 4 Preparing for Installation 5 Click OK Note If the voltage is not set correctly contact an authorized GE Service Representative Figure 3 14 System...

Page 92: ...ed 3 4 5 1 Video Format Confirmation 1 Turn ON the system and the check video transmission signal is set correctly 2 Press Config 3 From the System Configuration dialog box select the VCR ECG tab and...

Page 93: ...y images Check the grounding of the power cord and power outlet Be aware of RF sources Keep the unit at least 5m 16 4 ft away from other EMI sources Special shielding may be required to eliminate inte...

Page 94: ...CAUTION The Vivid 4 weighs 170 kg 375 lbs or more depending on installed peripherals when ready for use Care must be used when moving it or replacing its parts Failure to follow the precautions liste...

Page 95: ...formation for each peripheral device refer to the manufacturer s manual On board peripherals must be connected to one of the two available auxiliary power supplies on the right rear panel The total lo...

Page 96: ...ntry voltage 2 AUXILIARY AC OUTLET For use with external peripherals Voltages are set according to local country voltage If additional auxiliary outlets are required use the special cable provided by...

Page 97: ...Y C Video Out 4 pin connector for output to an S VHS VCR 3 S VIDEO IN Y C Video In 4 pin connector for input from an S VHS VCR 4 VIDEO OUT B W BNC connector for composite B W video output to a hard c...

Page 98: ...re 8 101 on page 115 3 After connecting the remaining peripherals and switching the system on configure the printer settings as described in the Printers Tab section on page 3 44 For more details abou...

Page 99: ...4 unit WARNING Whenever moving the Vivid 4 scanner the DeskJet Color Printer must be disconnected from the scanner DO NOT attempt to move the two units simultaneously without first disconnecting them...

Page 100: ...Gently push the connector in as far as possible 5 Rotate the locking latch 90 degrees clockwise to lock the connector into place vertical NOTE It is not necessary to turn OFF power to connect or disco...

Page 101: ...erent cable parts The main part trunk is a single cable connecting to the unit at one end and providing a cable splitter device at the other end The splitter contains five receptacles only three of wh...

Page 102: ...rds see Table 3 16 Figure 3 19 ECG Cable and Electrode Placement Table 3 16 ECG Cable Types Description Part No Full ECG Cable Kit AHA Americas Japan 2256477 Black wire LA 2269982 3 White wire RA 2269...

Page 103: ...e wall outlet plug should be connected to the fixed power socket that has a protective grounding conductor Never use an adapter or converter to connect with a power source plug for example a three pro...

Page 104: ...g the power cord connector into the AC input socket and secure it in place using the attached clip see Figure 3 20 4 Plug the other end of the power cord to the AC wall outlet Allow sufficient slack s...

Page 105: ...obe and application that were last used before the system was shut down If the probe has been removed since the system was last used the currently connected probes and their available applications are...

Page 106: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 3 Installation 3 35 Figure 3 21 Shut Down Options Screen...

Page 107: ...ght They can be adjusted using the Contrast and Brightness buttons on the front part of the display monitor All display monitor controls other than the contrast and brightness controls are factory adj...

Page 108: ...area set the date and time as follows Trackball to the Set button and press Select Use the alphanumeric keyboard to select the current date Select the format of the date and time display for example D...

Page 109: ...gure 3 24 Archive Tab Note The Archive tab will display even if the Archive Package option is not installed 3 Select Display patients gender on screen to display the patient s gender in the patient in...

Page 110: ...to which data is stored during backup from the Backup Device Is dropdown list 10 Select Eject MO Disk on Shutdown to have the system automatically eject the backup media when the unit is shut down 11...

Page 111: ...name is highlighted Trackball to the Change button and press Select The Font dialog box is displayed Trackball to the required font color style and size and press Select for each choice Trackball to...

Page 112: ...layed it will appear in the newly selected position until such time as this setting is changed 5 Activate the Duplicate Bodymark on Switch Window and or the Bodymark Menu on Freeze checkboxes as requi...

Page 113: ...xt to the name of the option being installed and use the alphanumeric keyboard to enter the password supplied by GE Note If an incorrect password is entered if the option has already been correctly in...

Page 114: ...43 Note If an option was not initially purchased or installed it may be acquired and or installed later During option installation type the relevant password in the appropriate Option window see Figu...

Page 115: ...with the Vivid 4 system Other generic printers are not approved and GE Medical Systems will not guarantee their correct operation on this system 3 In the Currently Installed Printers area trackball to...

Page 116: ...ions area complete the following as required Select the video system from the Video Mode dropdown list for example PAL or NTSC Select the SmartVCR option to enable Measurement and Analysis measurement...

Page 117: ...d define the default length of a cineloop when there is no ECG trace Select the Analyze only Systole checkbox to place markers only on the systole portion for synchronization and analysis purposes Sel...

Page 118: ...Tab RFI Configuration Monitoring should always be enabled this provides constant monitoring of critical parameters to ensure safe operation of the system Watchdog normally should not be enabled When...

Page 119: ...onfiguration Table 3 17 Recommended Vivid 4 Hardware Configurations RFI Models Item Vivid 4 BT03 ECG Second Re Image Port No board Front Board V7 3 TR V4 Ex Keyboard Revision 4 MUX Board V2 Sound Boar...

Page 120: ...hdog normally should not be enabled When enabled if there is a system lock up caused by the Back End this will turn OFF power at the Front End MOD Fan should be enabled only on BEPs based on Pentium I...

Page 121: ...ration Table 3 18 Recommended Vivid 4 Hardware Configurations RFT Models Item Vivid 4 BT03 ECG Second Re Image Port Revision E8 F8 Front Board V7 3 TR V4 Ex Keyboard Revision 4 MUX Board V2 Sound Boar...

Page 122: ...r 3 Installation 3 51 3 6 11 Technical Support History Tab The Technical Support History is where you would record any historical technical support information that may be relevant to a problem or any...

Page 123: ...station or a DICOM service using a TCP IP connection 3 7 2 Physical Connection 3 7 2 1 Ethernet Switch Connections An Ethernet Switch P N 066E0741 is used to connect the Vivid 4 system to an EchoPAC P...

Page 124: ...Workstation Figure 3 34 Hospital Network Connections See the EchoPAC PC Service Manual Part Number EP091298 for details about the EchoPAC PC and the required set up procedures EchoPAC PC Vivid7 Do no...

Page 125: ...s to the Hospital Information System HIS through HL7 protocol Users requiring this option should contact the GE representative 1 Press Config on the alphanumeric keyboard The System Configuration wind...

Page 126: ...L Chapter 3 Installation 3 55 4 Enable the DICOM printer option as required under the Services sub tab see Figure 3 36 Figure 3 36 Setting up DICOM Print 5 Enable any other remote server as required f...

Page 127: ...AL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 3 56 Section 3 7 Connectivity Setup Figure 3 37 Setting up Remote Server for MPEGvue 6 When done trackball to the OK button and press...

Page 128: ...d 2 Trackball to the Connectivity tab and press Select The Connectivity tab is displayed with the Views subsidiary tab selected by default as shown below Figure 3 38 Connectivity Tab 3 Trackball to th...

Page 129: ...omputer Name field shows the serial number of your Vivid 4 system 6 Enter the network information in the appropriate fields Leave the User Name as vivid3 Note The Default Gateway option is not always...

Page 130: ...window is displayed 3 Trackball to the Connectivity tab and press Select The Connectivity tab is displayed with the Views subsidiary tab selected by default as shown in Figure 3 38 on page 3 57 4 Trac...

Page 131: ...rage Server line in the Services table The Services tab is expanded as shown below Figure 3 43 Connectivity Tab Services for DICOM Server Expanded 10 Enter the AE Title and the Port No of the DICOM se...

Page 132: ...er 3 Installation 3 61 Figure 3 44 Connectivity Tab Settings for DICOM Print Expanded 12 Under Settings select the required Format Film Orientation etc see Figure 3 44 above 13 Proceed to the next sec...

Page 133: ...taflow configuration such as Int HD MOD 525 DICOM images are stored both on the internal hard drive and on media MOD The name DICOM is shown in parenthesis signifying that on the MOD the images are in...

Page 134: ...shed a red cross is displayed The properties of the DICOM storage server configuration are displayed on the right side of the tab for reference 6 If the connection was successfully established trackba...

Page 135: ...l Archive Int HD DICOM Server option from the drop down list at the bottom of the window If the system will not accept the dataflow or displays an error message there is an error in dataflow communica...

Page 136: ...reject any images that have a future date Ensure that the clock on the Vivid 4 ultrasound unit lags behind that of the EchoPAC system by a minute or two 1 Press Config on the alphanumeric keyboard Th...

Page 137: ...ver from the Name dropdown list The Services tab is updated as shown below Figure 3 48 Services Tab EchoPAC 8 Enter the IP address of the EchoPAC server in the IP address field 9 Select the Remote Arc...

Page 138: ...hive in the Placement of Report Files field OR If Image Vault Server is required make sure that the Image Vault Server checkbox is selected as shown in Figure 3 50 below then proceed as follows a Ente...

Page 139: ...mote Storage is displayed 5 Trackball to Remote Servers and press Select then trackball to the Check button and press Select If the connection is established a green checkmark is displayed If the conn...

Page 140: ...ew services to a dataflow In the event that any changes or additions are required contact the On line Center 3 7 6 2 Validating EchoPAC or Image Vault Operation 1 Press the Archive button on the contr...

Page 141: ...m is completely shut down Power disconnection before full system shutdown may cause system boot up failure This may require system recovery by booting up the system using the software CD Before storin...

Page 142: ...an ten degrees to prevent the unit from tipping over Once the destination is reached lock the front wheel brakes 3 8 4 Wooden Shipping Crate and Packaging Materials Verify that the following packing m...

Page 143: ...other means Table 3 21 Packaging Materials for the Cardboard Shipping Crate Item Description Qty 1 Press on envelope 2 2 Product Label 1 3 Input Pwr Cable 1 4 Platform wooden base 1 5 Ramp 1 6 Antista...

Page 144: ...al and any other manuals in an antistatic bag item 4 11 Lay the case base on the floor with the two wood strips facing upwards 12 Cover the top side of the base with the antistatic sheet 2 2 m x 3m 7...

Page 145: ...d appear next to each other in one corner The top cover has also a red mark that should be facing downwards in the corner where the other three marks meet Note Always insert the short leg of a clip in...

Page 146: ...in place using eight clips item 15 11 Secure the hand lever clip remover tool item 23 to the outside of the wooden crate using a screw 12 Secure the box with two steel strip loops item 16 around the...

Page 147: ...in an antistatic bag 11 Lay the wooden platform item 4 on the floor Leave sufficient free space around the platform 12 Lay the ramp item 5 on the floor and attach it to the narrow side of the wooden...

Page 148: ...included 1 Ensure that the ultrasound unit was properly packed that the brake pedal is in the parking position and that the front stopper is locked and prevents the unit from moving on the wooden plat...

Page 149: ...uals Installation Manuals for the peripheral units must be kept as part of the original system documentation This will ensure that all relevant safety and user information is available during the oper...

Page 150: ...he installation Specific language versions of the Vivid 4 User Manual may also be available Check with your GE Sales Representative for availability During and after installation supplementary documen...

Page 151: ...nstallation procedure or as required during servicing and periodic maintenance of the Vivid 4 ultrasound scanner Note Refer to Chapter 7 Diagnostics Troubleshooting for detailed descriptions of some o...

Page 152: ...Listen to fan spinning sound check filter go to monitoring and verify that all fans are working 2 Connect each of the probes available for the system Verify that the system properly recognizes each o...

Page 153: ...lider is moving on the Diagnostic panel 8 Perform an automatic lights test as described in the Automatic Lights Test on page 7 75 Verify that all the key LEDs are illuminated one by one 9 Select the S...

Page 154: ...able and has similar grey levels to that of the reference image Table 4 7 VCR Test Step Task Expected Result s 1 Connect a Sony 9600 VCR to the system restart the system and load an SVHS cassette Veri...

Page 155: ...image See B W Printer Tests below The image is displayed correctly 5 Connect the monitor to the Color Printer RGB connectors and check the image See Color Printer Tests below The image is displayed co...

Page 156: ...t you do not apply too much force to move the console and that the console movement is smooth 4 With the front handle still in the fully pressed position push the console to its minimum height Ensure...

Page 157: ...nd electrode pads 2 Perform the ECG diagnostic test as described in the ECG Phono Diagnostic Test on page 7 72 Verify that a single clear QRS signal is displayed Table 4 14 Phono Test Step Task Expect...

Page 158: ...y is displayed 8 Select the GetVolInfo button The message Media in drive G has following properties Media Label MY_LABEL Media File System Name FAT is displayed 9 Select the Get BTPsector button The m...

Page 159: ...abel Media File System Name CDUDF is displayed 8 Press the Get BTPsector button and respond OK to the message The message The Bytes Per Sector value of Media in Drive H is 2048 is displayed 9 Press th...

Page 160: ...he system remains in Standby mode indicated by the blinking On Off button LED 3 Turn ON the system s circuit breaker Press the On Off Button The system returns quickly to the Power ON state less than...

Page 161: ...effects 4 4 1 3S Probe Image Quality Tests 4 4 1 1 2D Noise Uniformity Test 4 4 1 2 2D Performance Test Table 4 18 2D Noise Uniformity Test 3S Step Task Expected Result s 1 With a 3S probe in the air...

Page 162: ...the phantom at position 3 Gain 37 Power 0 dB Depth 20cm Focus Max depth Dynamic Range 65dB 4 Record the maximum depth at which tissue can be differentiated from the noise The depth should be greater t...

Page 163: ...he Electronic Flow Phantom RMI1425A 2 Place the 3S probe above the vertical nails line position 1 3 Select a Cardiac preset and activate CFM 4 Set the following CFM parameters Gain Max possible withou...

Page 164: ...t and set the 2D Gain to 40 2 Activate CFM and set the following CFM parameters Gain Max possible without color noise Frame Rate 2 LVR MIN setting CFM Artifacts Test 3 Place the 3S probe on the RMI142...

Page 165: ...the following 2D parameters 2D Gain 50 2D Dynamic Range 60 2D Reject 10 2 Activate CFM and set the CFM Gain to the maximum possible without color noise 3 Activate M Mode Color M Mode CMM is activated...

Page 166: ...area is dark and that there are no flash strikes or other artifacts 5 Move the sample volume to different locations along the vessel Ensure that the flow looks clear the noise area is dark and that t...

Page 167: ...oise area is dark and that there are no flash strikes or other artifacts 5 Move the CW cursor to different locations along the vessel Ensure that the flow looks clear the noise area is dark and that t...

Page 168: ...hat the Doppler audio and display noise are homogenous and constant 3 Move the cursor angle multiple times from one half of the screen to the other Check that the noise level does not increase after t...

Page 169: ...parameters Depth 22cm Frame Rate 2 single focus Focus Location 12cm 3 Look for artifacts in the image Table 4 31 CFM Artifacts Test C358 Step Task Expected Result s 1 Use the Electronic Flow Phantom...

Page 170: ...w looks clear the noise area is dark and that there are no flash strikes or other artifacts 5 Move the sample volume to different locations along the vessel Ensure that the flow looks clear the noise...

Page 171: ...options 2 Set the following parameters 2D Gain Max Depth 2cm Gain Sliders Max Up to 5mm of white carpet noise should be displayed as shown in the left figure above Ensure that there are no white lines...

Page 172: ...the Electronic Flow Phantom RMI1425A 2 Place the 739L probe above the vertical nails line position 1 3 Select a Cardiac preset and activate CFM 4 Set the following CFM parameters Gain Max possible wi...

Page 173: ...eration Tests 739L Step Task Expected Result s 1 Select a Cardiac preset and set the 2D Gain to 40 2 Activate CFM and set the following CFM parameters Gain Max possible without color noise Frame Rate...

Page 174: ...w looks clear the noise area is dark and that there are no flash strikes or other artifacts 5 Move the sample volume to different locations along the vessel Ensure that the flow looks clear the noise...

Page 175: ...nter of the screen as shown opposite and write down the Gain value 4 Increase the Gain until background noise is displayed and write down the Gain value Ensure that the Gain difference is not higher t...

Page 176: ...4 6 show example configurations and should be used for reference only as each site has its own specific settings 4 5 1 Software Configuration Checks Table 4 42 Software Configuration Checklist Step T...

Page 177: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 4 Functional Checks 4 27 Figure 4 1 System Tab Configuration Example...

Page 178: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 4 28 Section 4 5 System Turnover Checklist Figure 4 2 System Options Tab Configuration Example...

Page 179: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 4 Functional Checks 4 29 Figure 4 3 Hospital Info Tab Configuration Example...

Page 180: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 4 30 Section 4 5 System Turnover Checklist Figure 4 4 VCR ECG Tab Configuration Example...

Page 181: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 4 Functional Checks 4 31 Figure 4 5 Archive Tab Configuration Example...

Page 182: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 4 32 Section 4 5 System Turnover Checklist Figure 4 6 Connectivity Local Settings Tab Configuration Example...

Page 183: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 4 Functional Checks 4 33 Section 4 6 Site Log Table 4 1 Vivid 4 Site Log Date Service Person Problem Comments...

Page 184: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 4 34 Section 4 6 Site Log Table 4 1 Vivid 4 Site Log Continued Date Service Person Problem Comments...

Page 185: ...his chapter contains block diagrams and functional explanations of the Vivid 4 electronic circuits Table 5 1 Contents in Chapter 5 Section Description Page Number 5 1 Overview 5 1 5 2 General Informat...

Page 186: ...beam forming system which can handle up to 192 element linear probes by use of multiplexing Signal flow travels from the Probe Connector Panel to the Front End FE Electronics to the BEP and finally is...

Page 187: ...ystem Block Diagram Vivid 4 BT03 Systems with RFI Configuration see Figure 5 1 below System Block Diagram Vivid 4 BT03 Systems with RFI Configuration see Figure 5 1 on page 5 3 System Cabling Diagram...

Page 188: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 5 4 Section 5 1 Overview Figure 5 2 System Block Diagram Vivid 4 BT03 Systems with RFT Configuration PCVIC...

Page 189: ...5 2252982 2258801 2252964 2253011 2252976 2252981 2253009 2253015 2292972 2276874 2300847 2253014 2276875 2269429 2253080 2253079 RS232 1 2266742 2269430 C B L F AC INPUT BOX 2252994 On off 2269430 AC...

Page 190: ...77 2252975 2252982 2258801 2252964 2253011 2252976 2252981 2253009 2253015 2292972 2276874 2300847 2253014 2276875 2253080 2253079 RS232 1 2266742 2269430 C B L F AC INPUT BOX 2252994 On off 2269430 A...

Page 191: ...ents as described in the following sections Front Board Assembly FB on page 5 16 MUX Board on page 5 18 Beamformer Board BF on page 5 19 Radio Frequency Interface RFI Board on page 5 21 OR Front End C...

Page 192: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 5 8 Section 5 1 Overview Figure 5 5 Front End Crate Block Diagram RFI Configuration...

Page 193: ...I Radio Frequency Interface Board OR FEC Front End Controller Board RFT Radio Frequency Tissue Board IMP Image Port Front Board TR4 Boards MUX Board 64Ch Beam Former A D Board RFT Board Image Port Boa...

Page 194: ...it bursts are generated on the Front board FB initiated by the transmit trigger pulse TXTRIG_L The transmit trigger starts the Transmit Pulse Generators TPG on the FB each generating 16 transmit pulse...

Page 195: ...to 80 V HV1 another providing a voltage ranging from 0 to 40 V HV2 and the third outputting 80 V for the multiplexers in the linear probes HV1 and HV2 are programmable through a serial interface from...

Page 196: ...EC or RFI and the FEC or RFI sends a signal to the high voltage power supply HVPS to program one of the supplies HVPS1 or HVPS2 Once the signal is transmitted it is received by the transmit pulsor TP...

Page 197: ...ed delay between the elements and immediately after this are received all the time aiming at one point at a specific depth In this way echo signals are collected from all points along a certain line f...

Page 198: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 5 14 Section 5 1 Overview Figure 5 8 Received Signal Path Block Diagram Systems with RFI Configuration...

Page 199: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 5 Components and Function Theory 5 15 Figure 5 9 Received Signal Path Block Diagram Systems with RFT Configuration...

Page 200: ...neral Description The FB Board has 64 identical receive channels The signal in each channel is fed via a Transmit Receive T R switch to a preamplifier The next step is a Time Controlled Gain TGC ampli...

Page 201: ...nels Each channel has two pulsars for different transmitted Tx voltages and a low signal receiver amplifier Both are connected to the same input output line The receiver is buffered from the transmitt...

Page 202: ...em to the BF Boards Expanding the low high channels in CW mode TEE motion control Monitoring temperature and voltage measurements via the I C FE Bus switch To reduce the noise in the FE cards the MUX...

Page 203: ...tput from the other BA Level 2 as shown in Figure 5 13 on page 5 20 These ASICs will add all the input and output channels with appropriate delay to give optimal receive focusing and beam steering as...

Page 204: ...1 Overview 5 4 4 2 Description of Operation Figure 5 13 Beamformer Board 64 Channel Block Diagram ADC ADC ADC ADC From prev board To next board BA2 FE bus 21 21 12 Bus Ifc RX N RX N 1 RX N 2 RX N 3 4...

Page 205: ...y the RFI board System Master Clock generator Transmitter Control ATGC Control Receive Focusing Control Analog Test Signal Generator High Voltage Control and supervision Probe Sensing Control of MUX b...

Page 206: ...clipping is performed 5 4 5 6 Color Flow Processing No Color Flow processing is performed here except for the RF demodulation previously mentioned The data used for Color Flow is tagged with the appr...

Page 207: ...ollows TX Power Supply 1 TX1 generates voltage levels for pulsed TX operations TX Power Supply 2 TX2 generates voltage levels for both pulsed and CW operations The RFI sets the voltage level for the t...

Page 208: ...A N A 9 2 5 V voltage Locally generated from 3 3 V 10 DSP CLOCK 50 MHz 11 15 V MONITOR voltage 12 EMIFA CLOCK 133 MHz 13 DSP_BSOE3_L signal 14 1 4 V voltage Locally generated from 3 3 V 15 OPTIONAL E...

Page 209: ...voltage 35 5 V voltage 36 5 V voltage 37 N A 38 N A 39 2 0 V voltage 40 FAN VENT 1 voltage 41 FAN VENT 2 voltage 42 10V_Ref_voltage 43 5V_ref voltage 44 5V_ref voltage 45 AUDIO RIGHT CHANNEL OUT 1 V...

Page 210: ...ter control ATCG control Receive focusing control Analog test signal control Probe sensing Transmit voltage control Control of Front Board switching Temperature sensing for the TEE probes Memory refre...

Page 211: ...Bus 5 4 7 1 Control The RFT Board has a local DSP with an external EPROM The DSP performs the following control tasks Handles communication with the BE over the VME Bus and the PCI Bus Updates filter...

Page 212: ...elow The IMP is used as a connection to the BEP All the instructions to the Front End Crate are sent via the IMP and all the outgoing data is transferred to the BEP via the IMP through the PCI Bus 5 4...

Page 213: ...PC2IP on page 5 38 Plug and Scan Card and Battery on page 5 38 Network Card Onboard on page 5 38 SCSI Card on page 5 38 Floppy Drive on page 5 39 Hard Disk on page 5 39 Magneto Optical Drive MOD on p...

Page 214: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 5 30 Section 5 1 Overview Figure 5 16 Back End Component Locations Vivid 4 BT03 Systems with RFI Configuration...

Page 215: ...dio Bracket PCI2IP Card Multifunction I O Card Key Board Control Card MO Driver ECG Module CD R Driver Floppy Driver HD Driver Mother Board P S Battery Slot 1 Slot 2 Slot 3 Slot 4 Slot 5 Slot 6 AGP Sl...

Page 216: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 5 32 Section 5 1 Overview Figure 5 18 Back End Block Diagram Vivid 4 BT03 Systems with RFI Configuration...

Page 217: ...ug Scan Card SCSI Card Network Card PCI2IP Card Empty slot Key Board Control Card MO Driver ECG Module CD R Driver Floppy Driver HD Driver P S Battery Video Out Mother Board S VGA To Front End To LAN...

Page 218: ...5 2 Central Processing Unit CPU The CPU controls and processes the internal Back End Processor operations In addition the CPU supports the Front End Crate via the IP card The CPU used in the Vivid 4 u...

Page 219: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 5 Components and Function Theory 5 35 Figure 5 20 BEP Motherboard Layout Vivid 4 BT03 Note Not drawn to scale...

Page 220: ...al Port Connector Header CPU Fan CPU Fan Header DIMM1 2 Memory DIMM Slots Ethernet LAN Ethernet Port Game Port Game Port J1a a P4SGA only 4xAGP J5 Floppy Disk Drive Connector J6 J7 IDE 1 2 Hard Disk D...

Page 221: ...10 seconds This is for emergencies only and should not be used under regular circumstances Standby Mode which enables system boot up in 30 seconds this option is enabled if the Plug Scan option is ins...

Page 222: ...PU The PCI2IP is sometimes referred to as the PCI Bridge 5 5 7 Plug and Scan Card and Battery The Plug Scan enables the Standby Mode during Plug and Scan operations and performs a legal shutdown of th...

Page 223: ...the Vivid 4 ultrasound unit The Hard Disk is controlled by the CPU via signals from the IDE Bus The Hard Disk is partitioned into four drives each of which is designated for different operations as fo...

Page 224: ...d to back up images and reports in the similar way as the MOD In addition the CD is used as the main source of software upgrades and other service utility operations The CDRW can write to 24X speed CD...

Page 225: ...5 5 15 Modem The Modem is a universal modem controlled by the CPU via the USB port The modem is used for iLinq InSite remote servicing This is a standard MultiTech modem that connects to an analog pho...

Page 226: ...image is captured and transmitted in RGB format to the S Video C Video Converter circuit and to the RGB to Color Printer Converter circuit The PC2IP board controls the PC VIC Power on and also contro...

Page 227: ...Timing Interlaced Vertical Start VS 53 Lines field Active Lines VA Lines frame PAL standard Total Lines VT 625 Lines frame PAL standard Frame Rate 25Hz PAL standard Table 5 6 NTSC Video Specification...

Page 228: ...ch receives and transmits Y C video signals is used for recording and re playing exams The appropriate type of VCR should be used for the local video signal PAL or NTSC 5 6 1 2 Black White Video Print...

Page 229: ...components and converted into DC voltages in the BEP and the Front End to support their operation Table 5 7 below shows which voltages are used in the various system components both in the BEP and the...

Page 230: ...e AC System see Figure 5 26 comprises the following components AC Isolation Transformer AC Distribution Box as described in AC Distribution Box on page 5 47 AC Input Box see AC Input Box on page 5 47...

Page 231: ...yback video signal from the VCR to the Image Port 5 7 3 1 AC Input Box The AC Voltage Input cord is connected to the AC Input Box It contains the Dual Rating Circuit Breaker 6A 200 240V or 12A 100 120...

Page 232: ...with specific requirements as indicated on the label near the AC input connector The AC transformer s primary and secondary voltages are set up by using the appropriate plugs connected in Jumper1 and...

Page 233: ...RECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 5 Components and Function Theory 5 49 5 7 4 Front End DC Power Distribution Figure 5 28 DC Distribution Front End Crate Vivid 4 Systems wi...

Page 234: ...PMX85N 80 80 40 40 5 10 PMX85P PMX85N DOLPHIN Front Bd MUX Bd BF2 BF1 FEC RFT Image Port J11 J21 J22 J14 J24 J25 J15 J16 J26 J33 J32 J31 J41 J42 J43 J53 J63 J73 J62 J52 J51 J61 J71 J72 125 125 125 125...

Page 235: ...nt End itself Air is sucked from the inside of the system through the Filter Cover and the Filter It passes the Fan Assembly directly into the upper part of the Front End where it forms a cooling air...

Page 236: ...th iLinq The following sections contain a brief introduction of iLinq s features 5 9 2 1 Web Server Browser The Service platform and other Service software use the iLinq web server and browser 5 9 2 2...

Page 237: ...service software across all GEMS products 5 9 3 1 Internationalization The user interfaces provided by the service platform are designed for GE personnel and as such are in English only At this time...

Page 238: ...Site use Every access request whether successful or not will be logged into a service access log that is viewable to authorized users See Chapter 7 Diagnostics Troubleshooting for more info about InSi...

Page 239: ...iew 6 1 1 Purpose of Chapter 6 Table 6 1 Contents in Chapter 6 Section Description Page Number 6 1 Overview 6 1 6 2 Input AC Voltage Configuration 6 2 6 3 Front End Voltages and Signal Indicators 6 3...

Page 240: ...ccording to specific requirements as indicated on the label near the AC input connector The AC transformer s primary and secondary voltages are set up by using the appropriate plugs connected in Jumpe...

Page 241: ...e have red green and yellow LEDs which indicate the board s status as follows Red LED Lit An error has occurred on the board unless otherwise specified Yellow LED Lit Software defined Green LED Lit Th...

Page 242: ...Front End Crate LEDs RFT Configuration The LEDs for each of the boards in the Front End Crate are summarized in the following subsections The LEDs are listed as viewed from the top of the board when...

Page 243: ...Transmit data to BEP During Processing LD502 Red Not Used LD501 Yellow Not Used LD500 Green Not Used LD12 Green 5 0 V status LD22 Green 2 0 V status LD11 Green 1 4 V status LD10 Green 1 5 V status LD...

Page 244: ...No Probe Lit if no probe is selected Yellow HV Over Range Lit if excessive HV is output possibly due to HV surveillance Table 6 6 RF and Tissue Processor LEDs LED Indicates When Lit Comments Red Board...

Page 245: ...Normally lit Green 5V PS is OK Normally lit Table 6 9 Front Board Assembly LEDs LED Indicates When Lit Comments Green 15V LVPS is OK Normally lit Green 15V LVPS is OK Normally lit Green AVee LVPS is O...

Page 246: ...Figure 6 3 Back End Power Supply Voltages Section 6 5 VIC Video Signal Setting The Vivid 4 ultrasound scanner and VIC may be configured to operate with either PAL or NTSC video systems as required 6 5...

Page 247: ...may be accessed directly from the three adjustment buttons located on the front of the monitor as shown in Figure 6 4 below NOTE All display monitor controls other than the contrast and brightness con...

Page 248: ...er to the recommended settings in Table 6 10 below 3 When the Contrast setting is correct press the center button again to make adjustments to the Brightness The Brightness adjustment bar is displayed...

Page 249: ...ar scanning mode enables you to perform three types of calibration as described in the following sections Monitor Calibration section page 6 13 Beamformer Calibration section page 6 14 Video Grabbing...

Page 250: ...DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 6 12 Section 6 7 Calibration 2 Trackball to the Calibration button and press Select The following calibration options are displayed Figure 6 6 C...

Page 251: ...to the Display option and press Select The Calibration Display window is displayed as shown below Figure 6 7 Calibration Display Window 3 Adjust the brightness of the monitor so that the black square...

Page 252: ...sk perform the BF calibration procedure 1 Access the calibration options as described in the Calibration section page 6 11 2 Trackball to the Calibration button and press Select The bias voltages for...

Page 253: ...lmost the same quality as the left side Note The video signal path is from the Back End SVGA board output to the VIC via the rear panel connector IP and then back to the Back End via the PC2IP cable 5...

Page 254: ...nd run diagnostic tools to locate system problems and failures The Vivid 4 ultrasound unit includes built in diagnostic tools that can be accessed from the scanner application Table 7 1 Contents in Ch...

Page 255: ...emperatures in the Front End motherboard and in the AC Distribution Box To check the Front End parts the tests must follow a logical sequence in order to identify the faulty component For example test...

Page 256: ...tools are accessed from 2D Mode by simultaneously pressing Ctrl F11 on the alphanumeric keyboard The Diagnostic menu is displayed as shown below Figure 7 1 Diagnostic Menu Note Once the test has been...

Page 257: ...ing Front End FE Diagnostics section on page 7 5 Tests various back end hardware components as described in the Performing Back End Diagnostics on the System section on page 7 69 Monitors all low volt...

Page 258: ...Diagnostic Test as described in the Beamformer BF Diagnostic Tests section on page 7 21 At this stage to continue to the Front End diagnostic tests it is necessary to re boot the system 3 Front Board...

Page 259: ...manual options see Manual Options on page 7 8 7 3 1 1 Auto Sequence 1 In the Diagnostic menu trackball to the Auto Sequence button and press Select The Automatic FE Testing dialog box is displayed as...

Page 260: ...under Execute Sequence and press Select if required 6 To resume testing from the point where you paused trackball to the Start button and press Select 7 At any stage during the testing sequence trackb...

Page 261: ...Diagnostics 7 3 1 2 Manual Options 1 In the Diagnostic menu trackball to the Front End button and press Select The FE Diagnostics dialog box is displayed as shown in Figure 7 3 or Figure 7 4 below Fig...

Page 262: ...DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 9 Figure 7 4 FE Diagnostics Dialog Box RFT Configuration Data Flow Map Test Description Area Test Area...

Page 263: ...splays a description of the selected diagnostic test Test Area Displays the buttons that enable you to perform diagnostic tests Each button has an LED that indicates the following by its color Green T...

Page 264: ...may be tested by performing either of the two tests Self Test or Full Test as required While the Full Test is in progress it may be necessary to wait a while since this requires longer than the Self T...

Page 265: ...s contact your local GE Medical Systems field engineer for assistance If errors were identified but the results were marginal perform calibration as described in the Calibration section on page 6 11 a...

Page 266: ...gnostics dialog box as described in the Accessing the Front End Diagnostic Options section on page 7 6 2 In the FE Diagnostics dialog box trackball to the IMP tab and press Select OR Trackball to the...

Page 267: ...Comment area and proceed as follows If no problems were identified proceed to the VME test as described in the VME Bus VME Diagnostic Tests for RFT Configuration section on page 7 15 If no problems we...

Page 268: ...section on page 7 13 If the VME Bus test is performed without first performing this test and errors are identified you must perform the IMP test to ensure that the VME board is the source of the error...

Page 269: ...area and proceed as follows If no problems were identified proceed to the RFT test as described in the RFT Diagnostic Tests for RFT Configuration section on page 7 17 If no problems were identified bu...

Page 270: ...ibed in the Image Port IMP Diagnostic Tests for RFT Configuration section on page 7 13 and in the VME Bus VME Diagnostic Tests for RFT Configuration section on page 7 15 If the RFT test is performed w...

Page 271: ...ere identified during the performance of the test 5 Review the test results in the Comment area and proceed as follows If no problems were identified proceed to the FEC test as described in the Front...

Page 272: ...Temp Test Diagnoses the temperature threshold for TEE probes Pencil ID Test Checks the pencil ID circuit in the FEC Note Before performing any of the FEC tests perform the IMP VME Bus and RFT tests as...

Page 273: ...dently The relevant LED changes color as follows Yellow The test is in progress Green The test has been completed and no problems were identified Red Problems were identified during the performance of...

Page 274: ...the VME Bus test and the IMP test to ensure that the BF board is the source of the error NOTE When replacing the BF perform calibration as described in the Calibration section on page 6 11 NOTE IMPOR...

Page 275: ...N 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 7 22 Section 7 3 Performing Front End FE Diagnostics Figure 7 11 BF Tab FE Diagnostics Dialog Box RFI Configuration Data Flow Map Test Description Area...

Page 276: ...ital Test 1 independently as follows Trackball to the Digital Test 0 button and press Select When complete trackball to the Digital Test 1 button and press Select 4 In the BF tab verify that the Comme...

Page 277: ...lternatively review the information manually using the scroll arrows on the right of the Ultrasound Scope window to select a vector between 0 and 63 The selected vector is displayed in the Vector fiel...

Page 278: ...neer for assistance If errors were identified but the results were marginal perform calibration as described in the Calibration section on page 6 11 and then repeat the test If errors were identified...

Page 279: ...11 If the FB tests are performed without first performing these tests and errors are identified you must perform the following tests in the order listed to ensure that the FB is in fact the source of...

Page 280: ...CTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 27 Figure 7 14 F B Tab FE Diagnostics Dialog Box RFI Configuration Data Flow Map Test Description Area Test...

Page 281: ...Test Checks the transmit switches as described in the TR Switch Test section on page 7 39 d FLA Jumper Test Checks the jumpers that connect the side elements to the center elements as described in th...

Page 282: ...test is in good condition 1 Connect either a 7S or 3S probe and lock it in place 2 Access the F B tab of the FE Diagnostics dialog box as described in the Accessing the Front Board Assembly Options se...

Page 283: ...he test result message in the Report dialog box If one or more channels were identified as bad trackball to the Save Report button and press Select then contact a GE Medical Systems service representa...

Page 284: ...esults it is imperative that only this probe is used and that it be in good condition 2 Access the F B tab of the FE Diagnostics dialog box as described in the Accessing the Front Board Assembly Optio...

Page 285: ...eckbox to display only bad channels in the Display area Mark bad channel Select this checkbox to manually mark the currently selected channel as bad 6 Check the test result message and proceed as foll...

Page 286: ...ard elements as follows Each element is represented by a button highlighted in green that displays the element and the channel number that is associated with that element Each button is highlighted ye...

Page 287: ...s Select and contact a GE Medical Systems service representative If no errors were identified proceed with the next test If no errors were identified but the problem persists contact your local GE Med...

Page 288: ...t is required if the tests are not being performed in the sequence described in the Performing Front End FE Diagnostics section on page 7 5 1 Disconnect all the probes 2 Access the F B tab of the FE D...

Page 289: ...Show Only Bad Channels Select this checkbox to display only bad channels in the Display area Mark bad channel Select this checkbox to manually mark the currently selected channel as bad 5 Check the t...

Page 290: ...p consists of four TR4 sockets each socket containing four vectors channels as follows Each vector is represented by a button highlighted in green that displays the number of the socket to which the T...

Page 291: ...s Select and contact a GE Medical Systems service representative If no errors were identified proceed with the next test If no errors were identified but the problem persists contact your local GE Med...

Page 292: ...B tab of the FE Diagnostics dialog box as described in the Accessing the Front Board Assembly Options section on page 7 26 3 In the Select Socket area of the F B tab trackball to the socket to which t...

Page 293: ...els were identified as bad trackball to the Save Report button and press Select then contact a GE Medical Systems Service Representative If no channels were identified as bad mark the channels suspect...

Page 294: ...has two phases namely the open pencil probe test and the close pencil probe test 1 Connect a 2D pencil probe 2 Access the F B tab of the FE Diagnostics dialog box as described in the Accessing the Fr...

Page 295: ...channels in the Display area Mark bad Channel Select this checkbox to manually mark the currently selected channel as bad 5 Check the test result message and proceed as follows If one or more relays...

Page 296: ...of the eight front board pencil probe relays as follows Each relay is represented by a button highlighted in green that displays the relay number Each button is highlighted in yellow when selected Eac...

Page 297: ...rt and press Select and contact a GE Medical Systems service representative If no errors were identified proceed with the next test If no errors were identified but the problem persists contact your l...

Page 298: ...are displayed in the Comments area 4 Check the test results and proceed as follows If no errors were identified proceed with the next test If no errors were identified but the problem persists contac...

Page 299: ...m focal beam Note You must reboot the system after performing the RAM test 1 Access the F B tab of the FE Diagnostics dialog box as described in the Accessing the Front Board Assembly Options section...

Page 300: ...ists contact your local GE Medical Systems field engineer for assistance If errors were identified but the results were marginal perform calibration as described in the Calibration section on page 6 1...

Page 301: ...Options 1 Access the FE Diagnostics dialog box as described in the Accessing the Front End Diagnostic Options section on page 7 6 2 In the FE Diagnostics dialog box trackball to the MUX tab and press...

Page 302: ...CTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 49 Figure 7 27 MUX Tab FE Diagnostics Dialog Box RFT Configuration Data Flow Map Test Description Area Test...

Page 303: ...l on page 7 59 AC Power Input Test on page 7 61 LVPS Voltage Test on page 7 63 HVPS Voltage Test on page 7 64 Temp Test on page 7 65 Show History on page 7 66 7 3 9 2 Normal Channels Diagnostic Tests...

Page 304: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 51 Figure 7 28 Normal Channels Diagnostic Test...

Page 305: ...Bad Channels Select this checkbox to display only bad channels in the Display area Mark bad channel Select this checkbox to manually mark the currently selected channel as bad 4 Check the test result...

Page 306: ...group consists of four TR4 sockets each socket containing four vectors channels as follows Each vector is represented by a button highlighted in green that displays the number of the socket to which t...

Page 307: ...s Select and contact a GE Medical Systems service representative If no errors were identified proceed with the next test If no errors were identified but the problem persists contact your local GE Med...

Page 308: ...ting 7 55 7 3 9 3 Expanded Low Channels Diagnostic Tests Tests output signals in CW Mode only when the input channels are connected to the output channels If one channel is missing the output will be...

Page 309: ...orming Front End FE Diagnostics 7 3 9 4 Expanded High Channels Diagnostic Tests Tests the output signals from the input of the high channels If one channel is missing the output will be 6db The input...

Page 310: ...SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 57 7 3 9 5 Port Test The Port test is a general logic control test Figure 7 32 Port Test RFI Con...

Page 311: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 7 58 Section 7 3 Performing Front End FE Diagnostics Figure 7 33 Port Test RFT Configuration...

Page 312: ...ll as temperatures 7 3 9 6 1 TEE Probe Control The TEE probe control test checks the position setting of the probe as well as its normal operation 1 Access the MUX tab of the FB Diagnostics dialog box...

Page 313: ...ECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 7 60 Section 7 3 Performing Front End FE Diagnostics Figure 7 35 TEE Probe Test Dialog Box 5 Verify the test results and then trackball to the OK b...

Page 314: ...accuracy by percentage 1 Access the MUX tab of the FB Diagnostics dialog box as described in the Accessing the MUX Diagnostic Test Options section on page 7 48 2 Trackball to the Advanced board tests...

Page 315: ...nt End FE Diagnostics 4 Trackball to the AC Power Input button and press Select The AC Power Input dialog box is displayed Figure 7 37 AC Power Input Dialog Box 5 Trackball to the Show button and pres...

Page 316: ...X Diagnostic Test Options section on page 7 48 2 Trackball to the Advanced board tests button and press Select The Advanced System Tests options are displayed as shown in Figure 7 34 on page 7 59 3 Tr...

Page 317: ...ox as described in the Accessing the MUX Diagnostic Test Options section on page 7 48 2 Trackball to the Advanced board tests button and press Select The Advanced System Tests options are displayed as...

Page 318: ...section on page 7 48 2 Trackball to the Advanced board tests button and press Select The Advanced System Tests options are displayed as shown in Figure 7 34 on page 7 59 3 Trackball to the Monitor but...

Page 319: ...Advanced board tests button and press Select The Advanced System Tests options are displayed as shown in Figure 7 34 on page 7 59 3 Trackball to the Monitor button and press Select The View Power opti...

Page 320: ...the Front End boards software versions and hardware revisions 1 Access the FE Diagnostics dialog box as described in the Accessing the Front End Diagnostic Options section on page 7 6 2 Trackball to...

Page 321: ...n the FE Diagnostics dialog box enables you to generate a report of the current test and failure results 1 Access the FE Diagnostics dialog box as described in the Accessing the Front End Diagnostic O...

Page 322: ...End hardware components Note The Back End Diagnostic Tests may be performed independently regardless of sequence 7 4 1 Accessing the Back End Diagnostic Options 1 In 2D Mode press Ctrl F11 on the alph...

Page 323: ...tions as described in the ECG Phono Diagnostic Test section on page 7 72 External Keyboard Enables you to test the keyboard options and lights in different modes as described in the External Keyboard...

Page 324: ...rmed for each of the speakers at various PRFs and frequencies Left and right are as seen when standing behind the unit 3 In the User controls area trackball to the Right radio button and press Select...

Page 325: ...r to an appropriate ECG simulator In this case select the External ECG radio button in the TestECGDevice dialog box 2 Access the HW Diagnostics menu as described in the Accessing the Back End Diagnost...

Page 326: ...n page 7 69 3 Trackball to the ECG Phono button and press Select The TestECGDevice dialog box is displayed as shown in Figure 7 47 on page 7 72 4 In the Select trace area trackball to the Phono radio...

Page 327: ...iagnostic Test section on page 7 77 7 4 4 1 Manual External Keyboard Test 1 Access the HW Diagnostics menu as described in the Accessing the Back End Diagnostic Options section on page 7 69 2 Trackbal...

Page 328: ...ector on the front panel of the unit and press the right middle and left of the switch Verify that the numbers 68 69 and 70 accordingly are displayed on the info bar 11 When you are finished trackball...

Page 329: ...al radio button and press Select 4 In the KB Panel tab deselect the Select checkbox Verify that after the Select LED is off the alphanumeric keyboards LEDs are dimmed 5 When you are finished trackball...

Page 330: ...nality of the alphanumeric keyboard keys 1 Access the HW Diagnostics menu as described in the Accessing the Back End Diagnostic Options section on page 7 69 2 Trackball to the Keyboard button and pres...

Page 331: ...splayed as shown below Figure 7 52 Util_MO_CDR_App Dialog Box 3 Trackball to the MO radio button and press Select 4 Trackball to the IsMediaAccess button an press Select The message Media in drive G i...

Page 332: ...d 7 Select No The message Media is outside is displayed Verify that the media has actually been ejected 8 Trackball to the Load button and press Select The message Media is inside is displayed Verify...

Page 333: ...s check the memory usage 7 4 7 1 Benchmark Testing 1 Access the HW Diagnostics menu as described in the Accessing the Back End Diagnostic Options section on page 7 69 2 Trackball to the Computer Testi...

Page 334: ...h indicates current computer results are higher or equal to those indicated in the lower line 7 4 7 2 Check Disk Bus 1 Access the HW Diagnostics menu as described in the Accessing the Back End Diagnos...

Page 335: ...Bench In the information displayed make sure that the speed values displayed are greater than 23MB sec This test checks the speed of information transfer between the hard disk and IDE BUS Bus Bench In...

Page 336: ...t The system enters Standby mode The light on the ON OFF button on the control console blinks repeatedly 3 Disconnect the AC power cable from the mains wall power outlet 4 Check that the ON OFF button...

Page 337: ...allel or USB port on the Vivid 4 rear connectors panel 1 Turn ON the system by pressing the ON OFF button on the control console and re boot into Technical Support Mode using the T plug While re booti...

Page 338: ...Connectivity Provides basic connectivity between the scanner and the OnLine Center OLC or the field engineer in the field Configuration Provides the interface to configure the various iLinq parameter...

Page 339: ...al Support tab as shown in Figure 7 58 below Figure 7 58 Technical Support System Configuration 2 Trackball to OK and press Select The System Configuration window closes Note If iLinq was not found to...

Page 340: ...location select Remote from the Profile Name drop down list 5 Trackball to the Start Communicator button and press Select Wait a while until the Service Login dialog box is displayed as shown below Fi...

Page 341: ...ct the required tool or application as described in the following sections Error Logs Page section on page 7 90 Diagnostics Page section on page 7 98 Image Quality Page section on page 7 111 Calibrati...

Page 342: ...Logs Page When the Error Logs page is selected as shown in Figure 7 62 below the log viewing options described in the following sections are available Log Viewer section on page 7 91 Log Browser sect...

Page 343: ...e following high level requirements Simple filtering of the scanner log s with filtering capabilities as a function of login access permissions Log visibility by all services modes Multiple instances...

Page 344: ...he power logs including errors monitoring dates and values as shown below Figure 7 65 Power Logs Temperature Displays all the temperature logs including errors monitoring dates and values as shown bel...

Page 345: ...Power logs in graphical format Select Power or Temperature from the Log option on the Logs Viewer menu and then select Plot Log to display the results as shown below Figure 7 67 Plot Log Window Plot...

Page 346: ...sitive text that you want to filter from the logs or pages that you are currently viewing as shown below Figure 7 69 Search for Text in Log or Page 7 5 3 1 4 Filter The Filter option on the Logs Viewe...

Page 347: ...Viewer Select the Exit Log Viewer option to return to the Common Service Desktop Figure 7 71 Exit Option 7 5 3 2 Log Browser When Log Browser is selected in the left pane of the Error Logs page you ca...

Page 348: ...mmon Service Interface Each system re boot will generate a new log file each log file can be viewed by selecting the appropriate link as shown in the example in Figure 7 73 below Figure 7 73 Viewing a...

Page 349: ...m this screen Figure 7 73 i e when you are connected to the system remotely or you may save the file to a floppy disk on the system itself Only three log files may be reviewed in the field DiagErr Sys...

Page 350: ...teractive The user is required to perform an operation on the ultrasound unit in order for the test to be completed successfully This option is not applicable when used remotely Non Interactive The te...

Page 351: ...rey abort Execute Button Enables you to start a selected diagnostic test Once the diagnostic is running the Execute button becomes the Abort button enabling you to stop the diagnostic Loop Count Field...

Page 352: ...AL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 7 100 Section 7 5 Common Service Interface Reset Power Supply Enables you to reset the Front End power supply Figure 7 76 Reset Power...

Page 353: ...ON 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 101 Restart Scanner Software System Shutdown Enables you to remotely shutdown or restart the system Figure 7 77...

Page 354: ...51 100 REVISION 3 VIVID 4 SERVICE MANUAL 7 102 Section 7 5 Common Service Interface Save System State to Log This performs the same system operation as that initiated when pressing Ctrl F or Alt D Fig...

Page 355: ...d CPU String Operations Tests the proper operation of the string commands LODS MOVS SCAS and STOS with 8 16 and 32 bit operands If errors are detected the failed commands and operands are listed CPU L...

Page 356: ...ing these diagnostics will also look for parity errors and other exceptions If encountered these are added to the test log as errors Memory Pattern Test Uses several test patterns to test as much memo...

Page 357: ...nd video controller 7 5 4 3 6 Frame Grabber Test Testing for proper operation of the Frame Grabber Board is achieved by performing the Video Grabbing Test For details refer to the Video Grabbing Test...

Page 358: ...0 Monitor Test Combination Test Checks the monitor alignment color depth and monitor resolution Four cross hairs are placed in each corner of the screen as well as RGB spectrum arrays and horizontal a...

Page 359: ...s as those performed by pressing Ctrl F11 during regular operation can be conducted Both Non interactive see Figure 7 81 or Figure 7 82 and Interactive see Figure 7 83 tests may be performed However t...

Page 360: ...for RFI Configuration on page 7 11 OR for systems with RFT Configuration Image Port Image Port IMP Diagnostic Tests for RFT Configuration on page 7 13 VME Bus VME Bus VME Diagnostic Tests for RFT Con...

Page 361: ...DICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 109 Figure 7 83 Interactive Tests NOTE Interactive tests are performed on the Front Board...

Page 362: ...est refer to the appropriate section as follows Transmit Test see Transmit Test on page 7 29 FLA Jumper Test see FLA Jumper Switch on page 7 31 TR Switch Test see TR Switch Test on page 7 39 Note On t...

Page 363: ...as shown below you can verify and calibrate image quality Figure 7 84 Image Quality Page Note This page is not populated in this version 7 5 6 Calibration Page When the Calibration page is selected as...

Page 364: ...ge When the Configuration page is selected you can configure the GSUI InSite No changes should be made in this page without the help of GE Medical Systems Service Personnel as any modifications made b...

Page 365: ...size already utilized IP Configuration shows the IP configuration of the different remote devices such as the network or the modem Network Status shows the communication status of active connections...

Page 366: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 7 114 Section 7 5 Common Service Interface Figure 7 88 System Shutdown...

Page 367: ...tem Shutdown There are three options provided described on previous page also see Figure 7 88 Reset Database used to empty the database and reset a new one see Figure 7 89 below Figure 7 89 Reset Data...

Page 368: ...gs CAUTION THIS PROCESS IS IRREVERSIBLE Before proceeding make sure you have a complete backup of the current settings if they will be required in the future When selecting Clean Userdefs the followin...

Page 369: ...Select the Prodiags Browser under the Configuration Utility option The following message is displayed Wait a few minutes until the following screen is displayed Figure 7 91 ProDiags Browser Schedule...

Page 370: ...EVISION 3 VIVID 4 SERVICE MANUAL 7 118 Section 7 5 Common Service Interface Figure 7 92 ProDiags Browser Task The Task tab indicates all the tasks that are available for activation This also indicates...

Page 371: ...TEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 119 Figure 7 93 ProDiags Browser Time Slot The Time Slot tab indicates the time slots available for...

Page 372: ...L 7 120 Section 7 5 Common Service Interface Figure 7 94 ProDiags Browser Log and Result The Log Result tab indicates the logs and results of any tasks that have been performed At any of the above men...

Page 373: ...NUAL Chapter 7 Diagnostics Troubleshooting 7 121 7 5 8 3 2 File Transfer Configuration The File Transfer Configuration utility can be used to transfer files to the GE Service center Follow the on scre...

Page 374: ...terface 7 5 8 3 3 iLinq Configuration Utility When the iLinq Configuration Utility option is selected in the left pane of the Configuration page see Figure 7 96 below the InSite Interactive Platform C...

Page 375: ...3 VIVID 4 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 123 Click Accept and select the modem device as the communication device to the OLC in the Device Connection tab also shown below Figu...

Page 376: ...p information required for the InSite Checkout tab as shown below When all the settings have been entered click the CHECKOUT NOW button twice The OLC will attempt to communicate with your ultrasound u...

Page 377: ...under the Technical Support Tab will allow you to save the iLinq IP settings to the hard disk as described below iLinq IP Save Procedure 1 In the System Configuration window click on the Technical Su...

Page 378: ...n window under the Technical Support Tab refer to Figure 7 99 above trackball to iLinq Settings Restore and press Select A blank screen appears wait for a while until the phone and modem options messa...

Page 379: ...tails and ordering information is displayed as described in Chapter 9 Renewal Parts Figure 7 100 Replacement Page Note This page is not populated in this version 7 5 10 PM Page When the PM page is sel...

Page 380: ...ity logging you can also enter your own comments as well as screen captures pertaining to system functionality These comments are saved as part of the daily logger report 1 When an error occurs press...

Page 381: ...ou expand the dropdown list in the Store Daily History area as shown below By default the current date is displayed Figure 7 103 Failure Report Calendar 3 Insert an empty PC formatted diskette Note An...

Page 382: ...252982 2258801 2252964 2253011 2252976 2252981 2253009 2253015 2292972 2276874 2300847 2253014 2276875 2269429 2253080 2253079 RS232 1 2266742 2269430 C B L F AC INPUT BOX 2252994 On off 2269430 AC DI...

Page 383: ...53016 2252977 2252975 2252982 2258801 2252964 2253011 2252976 2252981 2253009 2253015 2292972 2276874 2300847 2253014 2276875 2253080 2253079 RS232 1 2266742 2269430 C B L F AC INPUT BOX 2252994 On of...

Page 384: ...his chapter provides replacement procedures for system parts Table 8 1 Contents in Chapter 8 Section Description Page Number 8 1 Overview 8 1 8 2 Cover Replacement Procedures 8 2 8 3 Control Console C...

Page 385: ...2 Vivid 4 Ultrasound Unit Covers Legend to Figure 8 1 Label Item Label Item 1 Right Side Cover 7 Control Console Lower Cover 2 Left Side Cover 8 Control Console Upper Cover Front 3 Front Cover 9 Cont...

Page 386: ...power unit turn the power OFF and disconnect the power cord CAUTION Before removing circuit boards turn the power OFF and wear the ESD wrist band strap 8 2 1 2 Order of Cover Removal Remove the side c...

Page 387: ...rs four in total an example of which is shown below To easily access the screws turn the wheels inwards by 90 degrees Figure 8 2 Wheel Arch Screw 2 Slide the right side cover down and outwards until i...

Page 388: ...e Vivid 4 ultrasound unit as described in Chapter 3 Installation 8 2 3 3 Front Cover Removal Procedure 1 Pull out and remove the air filter as shown below Figure 8 3 Air Filter 2 Remove the side cover...

Page 389: ...nit taking care to first insert its upper edge under the top cover 2 Fasten the front cover with the three screws at the edge of the top cover as shown in Figure 8 5 above 3 Fasten the left and right...

Page 390: ...that secure the rear cover in position Two screws are located on the sides of the rear cover as shown below The other two screws are located in the AC Input Box Unit as shown in Figure 8 7 below Figu...

Page 391: ...edures 8 2 5 2 Preparations Shut down the Vivid 4 ultrasound unit as described in Chapter 3 Installation 8 2 5 3 Connector Panels Cover Removal Procedure 1 Remove the two screws located one on either...

Page 392: ...ed in the Side Covers Replacement Procedures on page 8 4 2 Remove the rear cover as described in the Rear Cover Replacement Procedures on page 8 7 3 Raise the control console to its maximum height 4 R...

Page 393: ...lug any cables from below the rear panels except for those indicated in this procedure 8 Viewing the unit from behind there are two small metal plates at the back of the top cover each with two screws...

Page 394: ...f the gas spring cover shown below and lower it into the collar of the top cover Figure 8 14 Gas Spring Cover Screws CAUTION The remaining section of the gas spring cover cannot be removed at this sta...

Page 395: ...sten it with the two screws shown in Figure 8 14 on page 8 11 7 Gently fit the left and right connector panels back to their original position on the top cover and fasten each with its four screws 8 M...

Page 396: ...s described in Chapter 3 Installation 8 2 7 3 Gas Spring Cover Removal Procedure The procedure for removing the gas spring cover is identical to the procedure for removing the lower section of the top...

Page 397: ...console lower rear cover as described in the Control Console Lower Rear Cover Removal Procedure on page 8 19 2 Loosen the screws fastening the control console lower cover in position and lower the fr...

Page 398: ...rear of the upper front cover refer to Figure 8 15 to fasten the front cover to the upper rear cover and tighten the 2 screws 4 Return and fasten the 12 screws 6 on each side and 2 in the front as sho...

Page 399: ...bed in the Vivid 4 17 Monitor Removal on page 8 24 CAUTION The Vivid 4 17 monitor is heavy At least two people must be available to lift the monitor Attempts to move this unit by one person alone coul...

Page 400: ...urn the control console lower cover as described in the Control Console Lower Cover Installation Procedure on page 8 18 5 Remove the control console lower rear cover as described in the Control Consol...

Page 401: ...the Control Console Lower Rear Cover Removal Procedure on page 8 19 4 Remove the 16 screws 7 from each side and 2 from the front that fasten the lower cover in place as shown in Figure 8 17 below Figu...

Page 402: ...unit as described in Chapter 3 Installation 8 2 11 3 Control Console Lower Rear Cover Removal Procedure 1 Raise the control console to its maximum height 2 Remove the 4 screws that fasten the rear co...

Page 403: ...cribed in the Control Console Upper Cover Front Removal Procedure on page 8 14 3 Remove the two 8mm securing nuts two on each side and one nut located below the keyboard parallel to the hand rest loca...

Page 404: ...as described in the Control Console Upper Cover Rear Removal Procedure on page 8 16 3 Remove the eight Allen 8mm securing screws four on each side from the rear of the upper console 4 Remove the Rear...

Page 405: ...aker Replacement Procedure on page 8 34 8 3 1 Vivid 4 17 Monitor Rear Cover Replacement Procedure NOTE The procedures below describe replacement of the Vivid 4 Monitor Rear Cover only P N 2303931 8 3...

Page 406: ...e the Vivid 4 control console is lowered to its minimum height 2 Carefully lift the replacement Monitor Rear Cover into position on the monitor 3 Return and fasten each of the 4 screws previously remo...

Page 407: ...ures 8 3 2 2 Preparation Shut down the Vivid 4 ultrasound unit as described in Chapter 3 Installation 8 3 2 3 Vivid 4 17 Monitor Removal 1 Turn OFF the Vivid 4 main circuit breaker located at the rear...

Page 408: ...n place before you return the monitor to the control console 1 Using two people lift the monitor into position on the control console 2 Fasten the monitor securing bracket in position using the four s...

Page 409: ...rocedure on page 8 19 2 Remove the control console lower cover as described in the Control Console Lower Cover Removal Procedure on page 8 18 3 Remove the control console upper cover front as describe...

Page 410: ...r refer to Figure 8 25 on page 8 26 3 Return and fasten the two M4 screws to secure the keyboard matrix to the metal frame one on each of the left and right corners refer to Figure 8 84 on page 8 99 4...

Page 411: ...its neighboring keypad see Figure 8 26 below 2 Use the screw driver to gently lever off the defective keypad until it is completely loosened from its base 3 Lift off and remove the defective keypad NO...

Page 412: ...allation 8 3 5 3 Keycaps Removal Procedure 1 Carefully place a small flat screw driver between the ATGC slider keycap you wish to remove and the surface of the keyboard 2 Use the screw driver to gentl...

Page 413: ...nob replacement procedure 8 3 6 2 Preparation Shut down the Vivid 4 ultrasound unit as described in Chapter 3 Installation 8 3 6 3 Rotary Knob Removal Procedure 1 Carefully place a small flat screw dr...

Page 414: ...the curved surface facing towards the center of the hole as shown in Figure 8 29 below Figure 8 29 Tension Plate Fitted in Centre of Knob 2 Carefully place the new rotary knob into position on the se...

Page 415: ...n the Control Console Upper Cover Front Removal Procedure on page 8 14 4 Unplug the four keyboard cables located at the bottom of the keyboard see Figure 8 24 on page 8 26 5 Loosen and remove the two...

Page 416: ...e left and right corners see Figure 8 25 on page 8 26 5 Reconnect the four keyboard cables to the keyboard previously disconnected see Figure 8 24 on page 8 26 6 Return the control console upper cover...

Page 417: ...Console Upper Cover Front Removal Procedure on page 8 14 3 Remove the four speaker screws one from each corner of both of the speakers 4 Gently pull the speakers towards you 5 Disconnect the wires at...

Page 418: ...ECTION POINT LOCATED ON THE REAR OF THE SCANNER TO THE RIGHT OF THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT 8 4 1 2 Preparation Shut down the Vivi...

Page 419: ...frame guiding tracks 2 Use the plastic latches to firmly lock in the card in position so that the connectors are firmly placed in their sockets 3 Return the front door of the crate to its original po...

Page 420: ...CATED ON THE REAR OF THE SCANNER TO THE RIGHT OF THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT 8 4 2 2 Preparation Shut down the Vivid 4 ultrasound...

Page 421: ...et previously removed that secures the TR4 boards to its original location above the TR4 boards 3 Return and tighten the four screws two on each side to fasten the bracket in position and secure the T...

Page 422: ...p and disconnect the AC power supply cable from the bottom of the DC power supply 3 Disconnect the AC power supply cable that is connected to the Tx P S HV P S 4 Remove the four small screws located o...

Page 423: ...ove the DC power supply refer to the DC Power Supply Removal Procedure on page 8 39 3 Remove the two screws located at the top of the Tx power supply unit 4 Remove the Tx power supply unit by pulling...

Page 424: ...probe sockets 3 Remove the right side cover as described in the Side Covers Removal Procedure on page 8 4 4 Remove the front door of the crate by unscrewing the two large screws at the top of the door...

Page 425: ...he front end crate back to its original position 5 Fasten the front end crate to the frame with the eight screws 6 Reconnect the PCI cable to the Image Port Board or RFI Board and slide it back to its...

Page 426: ...l Procedure 1 Remove the front end crate as described in the Front End Crate Removal Procedure on page 8 41 2 Unscrew and remove the four screws that secure the fans to the top of the front end crate...

Page 427: ...gure 8 38 on page 8 43 CAUTION Ensure that you position the fan in the correct airflow direction pushing air inwards into the crate and that you connect the fan to the connector that was marked during...

Page 428: ...paration Furnished Materials Confirm that the Back End Processor Kit P N 2315842 7 contains the following items Important The BEP arrives inside packing box P N 2328455 Make sure the returned BEP is s...

Page 429: ...only the following components ready installed CD R driver Floppy driver HD driver Cables and Brackets Memory 512 MB Therefore the FE will need to remove some components from the removed BEP and instal...

Page 430: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 8 Replacement Procedures 8 47 Figure 8 40 Back End Component Locations Systems with RFT Configuration...

Page 431: ...rchive 2 Press Utility and select Backup 3 Start full incremental backup onto the removable media 4 When the backup procedure is completed the yellow portion of the pie chart will be 0 8 5 2 3 Back En...

Page 432: ...KING THE NECESSARY ESD PRECAUTIONS 1 ALWAYS CONNECT YOURSELF VIA AN ARM WRIST STRAP TO THE ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER TO THE RIGHT OF THE POWER CONNECTOR 2 FOLLOW...

Page 433: ...in the MO Drive Installation Procedure on page 8 69 if applicable 2 Remove the Octopus Card Holder as described in the BEP2 Octopus Card Holder Removal Procedure on page 8 56 3 Install the SCSI Board...

Page 434: ...on page 8 51 13 Connect the VIC VGA OUT Left cable Part No 2252964 to the VGA OUT socket 1 on the BEP rear panel 14 Connect the VIC VGA IN Right cable Part No 2252965 to the VIC VGA IN socket 2 15 Co...

Page 435: ...cable Part No 2307289 to the Plug and Scan socket 16b and then connect the BEP Modem power cable Part No 2300847 to the power input socket 16a 31 Facing the side cover push the BEP as far to the right...

Page 436: ...Fan Set to ON checkbox activated UPS Change the UPS setting to Used or Not Used as appropriate If the Plug Scan option was purchased the UPS setting can be selected as Used If this option not purchas...

Page 437: ...cedure 1 Remove the two screws that secure the side cover in position from the rear of the BEP see Figure 8 44 below 2 Slide the cover backwards and remove it 8 5 3 4 BEP2 Side Cover Installation Proc...

Page 438: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 8 Replacement Procedures 8 55 Figure 8 45 Front Cover BEP2 Figure 8 46 Front Cover Removal BEP2...

Page 439: ...cover in position see Figure 8 45 on page 8 55 8 5 3 7 BEP2 Octopus Card Holder Removal Procedure 1 Remove the three screws that secure the Octopus Card Holder in position as shown in Figure 8 47 belo...

Page 440: ...al Procedure on page 8 56 5 Disconnect all cables connected to the Plug Scan board from the keyboard control card the battery and the power supply Note Counting from the top the board is positioned in...

Page 441: ...d unit as described in Chapter 3 Installation Voltage may exist in the system when a Plug Scan is present Ensure that a full shutdown of the system is performed otherwise voltage might be present that...

Page 442: ...the BEP Figure 8 50 Removing Plug Scan Velcro Strips 1 Place the battery inside the battery bracket Ensure that the cable is positioned at the lower end and that the rear of the battery with the two v...

Page 443: ...BEP 3 Remove the BEP side cover as described in the BEP2 Side Cover Removal Procedure on page 8 54 4 Remove the octopus card holder as described in the BEP2 Octopus Card Holder Removal Procedure on pa...

Page 444: ...Procedure on page 8 4 2 Disconnect the power cable from the rear of the BEP 3 Remove the BEP side covers as described in the BEP2 Side Cover Removal Procedure on page 8 54 4 If applicable remove the o...

Page 445: ...5 Back End Parts Replacement 7 Return the Vivid 4 rear and left and right side covers see the Rear Cover Installation Procedure on page 8 7 and the Side Covers Installation Procedure on page 8 4 Note...

Page 446: ...5 Disconnect the PCI cable that is connected to the BEP Note Counting from the top the PC2IP board is positioned in the fourth slot 6 Release the securing screw located on the left of the board bracke...

Page 447: ...val Procedure on page 8 56 5 Disconnect the frame grabber cable from the rear of the frame grabber board Note Counting from the top the frame grabber board is positioned in the fifth PCI slot 6 Releas...

Page 448: ...ables connected to the keyboard control board Note Counting from the top the keyboard control board is positioned in the sixth PCI slot 6 Release the securing screw located on the left of the board br...

Page 449: ...nstallation 8 5 11 3 CDRW Drive Removal Procedure 1 Remove the Vivid 4 left and right side covers and rear cover see the Rear Cover Removal Procedure on page 8 7 and the Side Covers Installation Proce...

Page 450: ...from the drive previously removed 2 Carefully slide the CDRW Drive into the front of the BEP refer to Figure 8 51 on page 8 66 3 Reconnect the cables data and power to the rear of the CDRW Drive 4 Ins...

Page 451: ...2 Remove the Vivid 4 left and right side covers refer to the Side Covers Removal Procedure on page 8 4 3 Remove the Vivid 4 rear cover refer to the Rear Cover Removal Procedure on page 8 7 4 Remove t...

Page 452: ...CSI card see SCSI Board Installation Procedure on page 8 61 4 Connect the data cable between the SCSI card and the rear of the MO Drive 5 Connect the power cable to the rear of the MO Drive Make sure...

Page 453: ...id 4 left and right side covers refer to the Side Covers Removal Procedure on page 8 4 3 Remove the Vivid 4 rear cover refer to the Rear Cover Removal Procedure on page 8 7 4 Disconnect all cables fro...

Page 454: ...le in position two from each side of the exposed BEP as shown in Figure 8 54 on page 8 71 9 Disconnect the footswitch cable from the KB Controller card 10 Disconnect the COM2 cable from the rear of th...

Page 455: ...COM 2 cable to the rear of the new ECG Module 5 Install the BEP left and right side covers as described in the BEP2 Side Cover Installation Procedure on page 8 54 6 Carefully return the BEP to its ori...

Page 456: ...und unit as described in Chapter 3 Installation 8 5 14 3 PC VIC Removal Procedure 1 Remove the left cover as described Side Covers Removal Procedure section page 8 4 2 Remove the BEP side cover as des...

Page 457: ...gure 8 57 above 3 Reconnect all the cables at the rear of the PC VIC module and the hard disk 4 Carefully slide the PC VIC into position in the same location as that of the removed PC VIC 5 Return and...

Page 458: ...dure 1 Disconnect the mains AC power cable for the mains power wall outlet 2 Remove the Vivid 4 left and right side covers refer to the Side Covers Removal Procedure on page 8 4 3 Remove the Vivid 4 r...

Page 459: ...he power supply in position on the BEP as shown in Figure 8 59 Note For the next steps it is recommended to carefully lie the BEP2 on its side to provide easier access 9 Cut any cable ties holding the...

Page 460: ...cables from the media drives MOD 2 and CD RW 3 13 Remove the screw that fastens the Ground cable and PC VIC in position as shown in Figure 8 62 below 14 Disconnect the power cable from the Hard Disk...

Page 461: ...BEP cutting any cable ties as necessary 19 Disconnect the mains power cable 8 from the motherboard beside the ECG location 20 Gently slide the power supply further downwards and then out of the BEP ca...

Page 462: ...oom and arrange each of the cables in the approximate position Carefully feed the cable connectors underneath the data cables 2 Holding the new BEP2 power supply in one hand connect the 4 pin external...

Page 463: ...ory boards etc are securely in position and that nothing has erroneously become disconnected during the power supply removal procedure 15 Stand the BEP2 upright 16 At the rear of the BEP2 return and l...

Page 464: ...d 4 ultrasound unit as described in Chapter 3 Installation 8 5 16 3 Hard Disk Removal Procedure 1 Remove the left cover as described Side Covers Removal Procedure section page 8 4 2 Remove the BEP sid...

Page 465: ...d disk make sure the jumper settings are as shown below 1 Place the new hard disk in position in the hard disk module holder 2 Reconnect all the cables at the rear of the hard disk 3 Fasten each of th...

Page 466: ...in the BEP2 Side Cover Installation Procedure on page 8 54 7 Return the Vivid 4 left cover as described in the Side Covers Installation Procedure on page 8 4 Note When a new hard disk arrives from the...

Page 467: ...6 1 3 AC Distribution Box Removal Procedure 1 Remove the Vivid 4 left right and rear covers as described in the Side Covers Removal Procedure on page 8 4 and Rear Cover Removal Procedure on page 8 7 2...

Page 468: ...n 2 Fasten the AC Distribution Box using the two screws located in the lower bracket of the AC Distribution Box 3 Tighten the screw that connects the Ground cable to the AC Distribution Box 4 Plug in...

Page 469: ...l Procedure on page 8 7 3 Unplug the internal power cable from the AC Distribution Box 4 Unscrew the Ground cable from the AC Distribution Box 5 Remove the two screws on both the left and the right si...

Page 470: ...ng all the cables at the back of the unit 4 Open the track by hooking a thumb or finger under the right side of one of the tracks and pulling the side of the track The track clicks open enabling you t...

Page 471: ...Allen keys as indicated in the cable replacement procedure 8 6 4 1 Preparation Shut down the Vivid 4 ultrasound unit as described in Chapter 3 Installation 8 6 4 2 AC BEP or FE Cable Removal Procedur...

Page 472: ...ide covers and the rear cover as described in the Side Covers Removal Procedure on page 8 4 and in the Rear Cover Removal Procedure on page 8 7 3 Remove the front end crate as described in the Front E...

Page 473: ...from the front of the unit and using an M4 Allen key loosen the two cap screws on the gas spring cable retaining bracket and slide the bracket forwards towards you sufficiently to release tension on...

Page 474: ...cable from the up down swivel handle by carefully sliding the cable end stopper out through the grooved slot on the handle assembly refer to the detail shown in Figure 8 88 on page 8 101 NOTE IMPORTA...

Page 475: ...sembly carefully aligning the groove on the cable with that of the groove in the slot refer to the detail shown in Figure 8 88 on page 8 101 7 Feed the gas spring cable down into the slot in the cable...

Page 476: ...16 Return the front end crate as described in the Front End Crate Installation Procedure on page 8 42 17 Return the Vivid 4 side covers and the rear cover as described in the Side Covers Installation...

Page 477: ...maximum height 2 Remove the control console lower rear cover as described in the Control Console Lower Rear Cover Removal Procedure on page 8 19 3 Remove the control console lower cover as described...

Page 478: ...cable Then slide the cable end stopper E and cable sleeve F out of the slot in the retaining bracket 9 Working from the front of the unit and using an M4 Allen key loosen the two cap screws on the sw...

Page 479: ...e cable end stopper out through the grooved slot on the handle assembly refer to the detail shown in Figure 8 88 on page 8 101 NOTE IMPORTANT Before finally removing the swivel release cable pay atten...

Page 480: ...79 on page 8 96 NOTE For clarification the letters appearing in square brackets in the following step correspond to the letters shown in Figure 8 77 on page 8 95 and Figure 8 78 on page 8 95 6 Workin...

Page 481: ...ribed in Chapter 3 Installation 8 6 7 3 Up Down Swivel Handle Removal Procedure 1 Remove the control console lower cover rear as described in the Control Console Lower Rear Cover Removal Procedure on...

Page 482: ...swivel release cable are now fully exposed providing easy access to complete the handle removal procedure 6 Using an M4 Allen key loosen the two cap screws on the cable retaining bracket and slide the...

Page 483: ...ease the swivel release cable from the up down swivel handle by carefully sliding the cable end stopper out through the grooved slot on the handle assembly refer to the detail shown in Figure 8 88 on...

Page 484: ...d of the swivel release cable from the vertical axis welding refer to the Swivel Release Cable Removal Procedure on page 8 94 11 Loosen and remove the M6 nut from the handle assembly axis screw see Fi...

Page 485: ...uring nut against the bracket to hold the cable firmly in position refer also to Figure 8 87 on page 8 100 6 Repeat step 4 to insert the swivel release cable end stopper into the grooved slot on the h...

Page 486: ...as described in the Front End Crate Removal Procedure on page 8 41 4 Lower the control console to its minimum height NOTE For clarification the letters appearing in square brackets in the following s...

Page 487: ...and then tighten the screw until the pin L at the lower end of the gas spring is loose enough to be removed and remove the pin NOTE For clarification the letters appearing in square brackets in the f...

Page 488: ...it the groove of the cable end sleeve C on the hinge F of the gas spring head 6 Turn the cable holder B over the cable end sleeve C and then fasten it with the securing screw A 7 Remove the Hex M8 scr...

Page 489: ...ved type Part No 2392789 with the same type of wheels refer to the following instructions Front Wheel Removal Procedure 2 on page 8 108 Front Wheel Installation Procedure 2 on page 8 110 8 6 9 1 Tools...

Page 490: ...s 8 107 Figure 8 94 Releasing the Spring Washer from the Center of the Transverse Arm 3 Remove the four bolts three wrench bolts and one Allen bolt that secure the front wheel to the chassis see Figur...

Page 491: ...ing the front and the brake in the left locked position preventing movement 1 Remove the Vivid 4 front cover as described in the Front Cover Removal Procedure on page 8 5 2 Remove the locking washer l...

Page 492: ...el 7 Proceed to step 1 of the Front Wheel Installation Procedure 2 on page 8 110 to install a new front wheel in place of the one removed 8 Repeat step 6 above to remove the other front wheel from the...

Page 493: ...lways remove and replace one wheel at a time before attempting to remove the second wheel 1 Having removed one wheel from the machine chassis as described in Front Wheel Removal Procedure 2 on page 8...

Page 494: ...ING When BOTH rear wheels are being removed the machine must be safely supported on two wooden blocks or on a jack capable of supporting the weight of the machine Always remove and replace one wheel a...

Page 495: ...remove and replace one wheel at a time before attempting to remove the second wheel 1 Having removed one rear wheel position the new wheel on the chassis in the same position as that of the wheel rem...

Page 496: ...the software CD 3 Hold down the ON OFF button for 5 seconds to shutdown the system The shutdown options are displayed on screen 4 Select Full Shutdown The system powers down 5 Hold down the ON OFF bu...

Page 497: ...priate flat and Phillips type screw drivers 8 8 1 2 Manpower One person 30 mins 8 8 1 3 Preparations 1 Shut down the Vivid 4 ultrasound unit as described in Chapter 3 Installation 2 Remove the Connect...

Page 498: ...r on the right side of the Vivid 4 ultrasound unit s top surface below the control console as shown in Figure 8 102 and next to the VCR if present allowing sufficient room to access the cables at the...

Page 499: ...ribed in the Connector Panels Cover Installation Procedure on page 8 8 7 Reconnect the mains AC power cable from the wall outlet 8 Turn ON the main circuit breaker located on rear of the Vivid 4 9 Tur...

Page 500: ...r as described in the Connector Panels Cover Removal Procedure on page 8 8 2 Release the B W printer by undoing the fastening belt secured around it and around the VCR if present 3 Disconnect the Remo...

Page 501: ...The VCR will not operate without initial setup and adjustments as described below Figure 8 105 Mitsubishi Video Cassette Recorder VCR 8 8 2 1 Tools No special tools needed 8 8 2 2 Manpower One person...

Page 502: ...cable P N 2269430 to the power outlet on the Vivid 4 rear panel and the other end to the AC power socket on the VCR rear panel see Figure 8 107 on page 8 119 3 Using the S video cable P N 2266743 con...

Page 503: ...ng the power switch on VCR front panel 6 Press Config then from the System Configuration window activate the VCR ECG tab 7 Under the VCR Options list choose Mitsubishi from the VCR model drop down lis...

Page 504: ...3 VIVID 4 SERVICE MANUAL Chapter 8 Replacement Procedures 8 121 8 From the Video Mode drop down list choose PAL or NTSC according to video system in your region then press OK to apply the changes 9 P...

Page 505: ...the Enter button to apply your selection To configure the Mitsubishi VCR 1 Press the VCR Play B button on the Vivid 3 control console to display the VCR menus on the Vivid 3 monitor 2 Using the Jog bu...

Page 506: ...igure 8 113 below then press the Enter button to continue to the next menu 5 Skip Menu3 Figure 8 114 by pressing the Enter button to continue to the next menu 6 Select the RS 232C settings according t...

Page 507: ...or and Audio switches are in the positions shown in Figure 8 117 below if necessary adjust accordingly then move the Key Lock switch to the ALL position to lock all keys 10 Perform the VCR Test as des...

Page 508: ...e 8 8 2 If optional shelf P N H45011CN is present refer to the Sony UP 21MD Color Video Printer Removal Procedure on page 8 138 3 Disconnect the RS 232C cable the Audio and S Video IN and OUT cables a...

Page 509: ...arations 1 Shut down the Vivid 4 ultrasound unit as described in Chapter 3 Installation 2 Remove the Connectors Panels Cover as described in the Connector Panels Cover Removal Procedure on page 8 8 3...

Page 510: ...d next to the B W Printer if present allowing sufficient room to access the cables at the rear of the VCR Figure 8 121 Video Cassette Recorder VCR Installed in the Vivid 4 Ultrasound Unit Figure 8 120...

Page 511: ...lled Figure 8 122 Sony VCR Connection Cables AUDIO IN cable From Left Connectors Panel To VCR Audio Input Out R Connector Red Out L Connector Black BLACK CH1 RED CH2 RED BLACK RED BLACK AUDIO OUT cabl...

Page 512: ...n page 3 45 Note Sony VCR Reset Counter If the Reset Counter button is activated while the VCR is in the Stop Pause or Record state the counter display is reset to zero 0 00 00 8 8 3 5 Sony VCR Remova...

Page 513: ...s necessary to order the Panasonic VCR RS 232 Adaptor Part No 066E8200 Figure 8 123 Panasonic Video Cassette Recorder VCR 8 8 4 1 Tools No special tools needed 8 8 4 2 Manpower One person 30 mins 8 8...

Page 514: ...mum height 2 Place the VCR on the Vivid 4 ultrasound unit s top surface left side below the control console as shown in Figure 8 125 below allowing sufficient room to access the cables at the rear of...

Page 515: ...onic VCR Connection Cables S VIDEO IN cable From Internal IO Module Connector A11 To S Video Input S VIDEO OUT cable From Internal IO Module Connector A10 To S Video Output Power cable From Distributi...

Page 516: ...ector Panels Cover Installation Procedure on page 8 8 9 Reconnect the mains AC power cable from the wall outlet 10 Turn ON the main circuit breaker located on rear of the Vivid 4 11 Turn ON the Vivid...

Page 517: ...Release the B W printer by undoing the fastening belt around it together with the VCR if present 3 Disconnect the RS 232C cable with adaptor the Audio and S Video IN and OUT cables and the Power cabl...

Page 518: ...r on the rear panel Figure 8 128 SONY UP 21MD Color Video Printer 8 8 5 1 Tools Use the appropriate flat and Phillips type screw drivers 8 8 5 2 Manpower One person 30 mins 8 8 5 3 Sony UP 21MD Color...

Page 519: ...side compartment as shown in Figure 8 129 below 6 Connect the Ground cable hanging out of the Bottom Metal Cover to the Ground screw at the rear of the UP 21MD Color Video Printer as shown in Figure 1...

Page 520: ...Metal Cover slot see Figure 8 129 on page 8 136 to the UP 21MD Color Video Printer as shown in the diagram in Figure 8 131 below Figure 8 131 Sony UP 21MD Color Video Printer Connection Cables R G B...

Page 521: ...Color Video Printer Installed on the Vivid 4 8 8 5 4 Sony UP 21MD Color Video Printer Removal Procedure 1 Shut down the Vivid 4 ultrasound unit as described in Chapter 3 Installation 2 Slide the UP 2...

Page 522: ...0 3 Printer Kit P N 2415435 contains the following items 8 8 6 1 Tools Use the appropriate flat and Phillips type screw drivers 8 8 6 2 Manpower One person 15 mins 8 8 6 3 Preparations 1 Shut down the...

Page 523: ...age Power IN socket on the rear of the printer as shown in Figure 8 134 below WARNING To avoid electrical hazards the HP 6540 3 USB Deskjet Color Printer must be connected to the AC Supply of the Vivi...

Page 524: ...click Cancel see Figure 8 135 below Figure 8 135 Found New Hardware Wizard 7 Insert the HP6540 3 driver installation CD P N 2415381 provided with the printer into the CD ROM drive 8 Browse the CD Dri...

Page 525: ...8 Peripherals 9 When the Add Printer Wizard opens click Next see Figure 8 138 10 When prompted verify that Local Printer is selected If not activate the Local printer radio button as shown in Figure...

Page 526: ...VICE MANUAL Chapter 8 Replacement Procedures 8 143 11 Click Next 12 From the displayed list of available ports choose one of the available USB ports as shown in Figure 8 140 13 Click Next Figure 8 139...

Page 527: ...pherals 14 From the list of Printers highlight the HP Deskjet 6500 Series see Figure 8 141 15 Click Next 16 When prompted choose Yes if you want this printer to be the default printer see Figure 8 142...

Page 528: ...es 8 145 18 When the following message appears verify that Do not share this printer is selected as shown in Figure 8 143 19 Click Next 20 In response to the Print a test page message choose No and th...

Page 529: ...rals The Add Printer Wizard displays the following message 21 Click Finish to close the Wizard The Printer Properties Dialog opens with the General Tab activated as shown in Figure 8 146 22 Click on P...

Page 530: ...message is displayed 23 Confirm that test page was printed properly then click OK The Printer Properties Dialog opens with the Ports Tab activated as shown in Figure 8 148 24 Confirm that a USB port...

Page 531: ...ripherals The Printer Properties Dialog opens with the Advanced Tab activated as shown in Figure 8 149 25 Click on Printing Defaults The Printing Defaults dialog box opens with the Printing Shortcuts...

Page 532: ...etc to your liking then click OK The following message is displayed 27 Press any key to reboot the system and then eject the CD from the CD tray 28 Press the Config button on control console and from...

Page 533: ...tly Installed Printers list then click OK 8 8 6 5 HP 6540 3 USB Deskjet Color Printer Removal Procedure 1 Turn OFF power to the USB printer 2 Unplug the AC power cable from the Power IN socket at the...

Page 534: ...n 9 1 Overview 9 1 1 Purpose of Chapter 9 This chapter gives you an overview of replacement parts for the Vivid 4 ultrasound unit Table 9 1 Contents in Chapter 9 Section Description Page Number 9 1 Ov...

Page 535: ...st of Abbreviations Assy Assembly BEP Back End Processor Ctrl Control FEP Front End Processor FRU 1 Replacement part available in parts hub FRU 2 Replacement part available from the manufacturer lead...

Page 536: ...1 Mechanical Hardware Parts Figure 9 1 Mechanical Hardware Parts Diagram 1 Table 9 2 Mechanical Hardware Parts Diagram 1 Item Part Name Part Number Description Qty FRU 101 Cover Rear 2258594 1 2 102 C...

Page 537: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 9 4 Section 9 3 Renewal Parts Lists and Diagrams Figure 9 2 Mechanical Hardware Parts Diagram 2 201 202 203 204 205 206 207...

Page 538: ...Part Name Part Number Description Qty FRU 201 Cover Right 2258600 1 1 202 Cover Top Lower Section 2258602 1 2 203 Control Console Upper Cover Front V4 2405642 1 1 204 Control Console Upper Cover Rear...

Page 539: ...3 Table 9 4 Mechanical Hardware Parts Diagram 3 Item Part Name Part Number Description Qty FRU 301 Filter Air 2301690 1 2 302 Cover Gas Spring Sleeve V4 2261575 2 1 2 303 Probe Cable Holder V4 226384...

Page 540: ...Diagram 4 Table 9 5 Mechanical Hardware Parts Diagram 4 Item Part Name Part Number Description Qty FRU 501 Wheel Castor Rear 2389102 Rear 2 1 502 Wheel Castor Front 2392789 Front with brake and trans...

Page 541: ...Illustrated Item Part Name Part Number Description Qty FRU 601 Swivel Release Cable 2392103 1 2 602 Monitor Service Cover 2392137 1 2 603 Screw Kit 2253052 1 1 604 Keycaps Kit V4 2394000 2 1 1 605 Ve...

Page 542: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 9 Renewal Parts 9 9 9 3 2 AC System Parts Figure 9 5 AC System Parts 701 702 703 704 705 706 707 708...

Page 543: ...reaker 2252994 1 1 702 AC Voltage Distribution Box 2399304 2 1 1 703 Connector AC Voltage Primary 100 120V 2253054 100 120V 1 2 704 Connector AC Voltage Primary 200 240V 2253056 200 240V 1 2 705 Conne...

Page 544: ...Front End Parts Diagram 1 Table 9 8 Front End Parts Diagram 1 Item Part Name Part Number Description Qty FRU 801 Front End Crate 2253030 2 including back plan MB 1 2 802 Front End Crate 2253030 3 syst...

Page 545: ...d Parts Diagram 2 Item Part Name Part Number Description Qty FRU 1001 PCB Front Board and TR4 Assembly 2253033 3 FB 3Con V7 1 1 1002 PCB TR4 Board 2418129 16 1 1003 PCB Multiplexer MUX 2253038 3 1 1 1...

Page 546: ...3 Item Part Name Part Number Description Qty FRU 1011 PCB RF Tissue RFT MLA2 FB200992 Applies to systems with RFT configuration only 1 1 1012 PCB Beamformer BF 64ch FB200900 1 1 1013 PCB Image Port 3...

Page 547: ...and Diagrams 9 3 4 Back End Parts Figure 9 9 Back End Parts Diagram 1 Table 9 11 Back End Parts Diagram 1 Item Part Name Part Number Description Qty FRU 1021 Keyboard Trackball 2274623 1 1 1022 Keybo...

Page 548: ...bly Spanish 2393953 4 1 2 Keyboard Overlay Assembly Portuguese 2393953 5 1 2 Keyboard Overlay Assembly Italian 2393953 6 1 2 Keyboard Overlay Assembly Russian 2393953 8 1 2 Keyboard Overlay Assembly D...

Page 549: ...attery 2300858 1 1 1052 PCB Plug Scan Card 2316279 1 1 1053 MO Drive 5 25 2277189 This hardware is an option that should be ordered P N H45011CF 1 1 1054 Footswitch 066E0007 1 2 1055 BEP Vivid 4 Assy...

Page 550: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 9 Renewal Parts 9 17 Figure 9 12 Back End Parts Diagram 4 1061 1062 1063 1064 1065 1066 1067 1068 1069...

Page 551: ...PC VIC 1 1 1062 Hard Disk Drive HDD 40 to 100 GB 2408479 1 1 1063 Floppy Drive 2346034 1 2 1064 CD Drive RW 2413449 1 1 1065 PCB SCSI Card 2271149 2 This hardware is an option that should be ordered...

Page 552: ...Cable VGA VIC PC 2252965 1 2 1103 Cable Input Box AC Distribution Box 2252966 1 2 1104 Cable Soft Shutdown 2252967 1 2 1105 Cable Ground VIC 2252968 1 2 1106 Cable Ground Back End 2252969 1 2 1107 Cab...

Page 553: ...k End Matrix Panel 2252978 1 2 1115 Cable ECG Back End Internal 2252979 1 2 1116 Cable Patient I O int ECG ECG Phono 2252980 1 2 1117 Cable ECG Int Footswitch Keyboard Control 2252982 1 2 1118 Cable C...

Page 554: ...C Box 2254089 1 2 1128 Cable ECG BEP Cable USA External 2256477 1 1129 Cable ECG BEP Cable Europe External 2256478 1 1 1130 Cable External ECG with Echo Stress Option 2256685 1 2 1131 Cable Internal A...

Page 555: ...S232 Adaptor for Panasonic VCR 066E8200 1 1 1138 Cable Video IN Rear Panel 2266743 1 2 1139 Cable Video OUT Rear Panel 2266744 1 2 1140 Cable Audio OUT Rear Panel 2266745 1 2 1141 Cable Audio IN Rear...

Page 556: ...tion Box 2292972 1 N 1150 Cable USB Cable 2297905 1 2 1151 Cable ECG Internal COM 2 2298089 1 N 1152 Cable Power Peripherals and Printer 2300857 1 2 1153 Cable Cable Power Asia 2269428 2 1 1 1154 Cabl...

Page 557: ...EVISION 3 VIVID 4 SERVICE MANUAL 9 24 Section 9 3 Renewal Parts Lists and Diagrams 9 3 6 Software Table 9 16 Software Item Part Name Part Number Description Qty FRU 1301 S W Kit contains s w for syste...

Page 558: ...08 Probe 5S Sector H40422LA H4901RA 1 1 1409 Probe 10L 739L Linear H40412LG 1 1 1410 Probe 39A 12L Linear H40412LH 1 1 1411 Biopsy Kit for 10L 12L E83885MM 1 N 1412 Probe 10S H4901PC 1 1 1413 Probe P5...

Page 559: ...MD3000E S VHS RS232C PAL H45011PT 1 2 1502 VCR Mitsubishi HS MD3000U S VHS RS232C NTSC H45011PN 1 2 1503 VCR Cassette S VHS H45001PG 1 N 1504 Color Printer Tray V4 2393556 1 2 1505 Sony UP 21MD PAL N...

Page 560: ...77 2252975 2252982 2258801 2252964 2253011 2252976 2252981 2253009 2253015 2292972 2276874 2300847 2253014 2276875 2269429 2253080 2253079 RS232 1 2266742 2269430 C B L F AC INPUT BOX 2252994 On off 2...

Page 561: ...16 2252977 2252975 2252982 2258801 2252964 2253011 2252976 2252981 2253009 2253015 2292972 2276874 2300847 2253014 2276875 2253080 2253079 RS232 1 2266742 2269430 C B L F AC INPUT BOX 2252994 On off 2...

Page 562: ...n the backplane the AC distribution box and the DC distribution box that could be dangerous Be sure to disconnect the system power plug and to open the main circuit breaker before you remove any parts...

Page 563: ...canner or nearby 10 2 2 Quality Assurance In order to gain accreditation from organizations such as the American College of Radiology USA it is the customer s responsibility to have a Quality Assuranc...

Page 564: ...nner as recommended in the Periodic Maintenance Schedule in order to retain the high levels of safety dependability and system performance Your GE Medical Systems Service Representative has an in dept...

Page 565: ...rective maintenance or as required by your facility s QA program Peripheral Current Leakage Checks Also after corrective maintenance or as required by your facility s QA program Surface Probe Current...

Page 566: ...abrasions Equipment for loose or missing hardware The keyboard for defects Brakes for defective operation DANGER To avoid electrical shock hazard do not remove the panels or covers from the machine T...

Page 567: ...le wrist strap 3M 3051 conductive Ground cord Anti Static Vacuum Cleaner 46 194427P278 46 194427P279 120V 230V Air Filter 2256176 Air intake Safety Analyzer 46 285652G1 DALE 600 KIT or equivalent for...

Page 568: ...if they have any problems or questions about the equipment 2 Paperwork Fill in the appropriate details at the top of the Vivid 4 Preventative Maintenance Inspection Certificate see page 10 31 Note al...

Page 569: ...Mode operation Check the basic system controls that affect this mode of operation 5 Applicable Software Optionsa a Some software may be considered as standard depending upon system model configuratio...

Page 570: ...ssary 3 Color Printer Verify hardcopy output of the Color video page printer Clean heads and covers if necessary 4 DICOM Verify that DICOM is functioning properly Send an image to a DICOM device 5 InS...

Page 571: ...lps to maintain a clean working environment Prior to cleaning turn OFF the power to the system Use a fluid detergent in warm water on a soft damp cloth to carefully wipe the entire system Be careful n...

Page 572: ...the air filter replacement part number Table 10 9 Air Filter Cleaning Step Item Description 1 Air Filter Remove the air filter located at the front of the scanner above the probe connector sockets Cl...

Page 573: ...MOD disks must be stored away from dust and cigarette smoke Do not use alcohol or benzene to clean the MOD cartridge 10 CD Drive Clean the drive head and media with the vendor supplied cleaning kit A...

Page 574: ...0 13 10 5 6 Diagnostic Checks Optional To complete the periodic maintenance checks perform the diagnostic tests as described in Chapter 7 Diagnostics Troubleshooting Review the system error log for an...

Page 575: ...heck of the probe pins and system sockets before plugging in a probe Avoid pulling pinching or kinking the probe cable since a damaged cable may compromise the electrical safety of the probe To avoid...

Page 576: ...TEE and Interoperative probes often have special usage considerations always refer to the individual probe manufacturers handling instructions user manual 10 6 4 Probe Cleaning 10 6 4 1 Basic Probe Cl...

Page 577: ...ogen free probe sheaths Probes used during neurological surgery must NOT be sterilized with liquid chemical sterilants because of the possibility of neurotoxic residue remaining on the probe 10 6 4 2...

Page 578: ...will receive and or open the package NOTE The US Department of Transportation DOT has ruled that items that were saturated and or dripping with human blood that are now caked with dried blood or whic...

Page 579: ...under test must not be connected to other electrical equipment Remove all interconnecting cables and wires The operator and patient must not come into contact with the unit while the tests are being...

Page 580: ...rmal Condition Open Ground Reverse Polarity Open Neutral Mains Applieda a Mains Applied is the Sink Leakage Test In this test mains supply voltage is supplied to the part or equipment to determine the...

Page 581: ...the Grounding protective earth conductor have been reversed If later tests indicate high current leakages this should be suspected as a possible cause and the outlet wiring should be visually inspecte...

Page 582: ...hese steps to test the Ground wire resistance 1 Turn the Vivid 4 unit OFF 2 Connect one of the Ohmmeter probes to the tested AC wall outlet Ground pin 3 Using the other Ohmmeter probe touch the expose...

Page 583: ...powered ON Be sure to turn the unit power OFF before switching them using the Polarity switch and or the Neutral switch 10 7 5 2 Generic Procedure The test verifies the isolation of the power line fro...

Page 584: ...le 10 14 Record all data on the Vivid 4 Preventative Maintenance Inspection Certificate see page 10 31 Table 10 17 Typical Data Sheet for Chassis Source Current Leakage Test Unit Power Tester Polarity...

Page 585: ...s should be made with the Ground Open and Closed with the Vivid 4 unit ON and OFF and with the power line polarity Normal and Reversed Record the highest reading For each combination the operating con...

Page 586: ...d Patient Lead Sink Leakage Isolation Test NOTE Refer to the procedure in the IEC 60601 1 CAUTION Line voltage is applied to the ECG leads during this test To avoid the risk of possible electric shock...

Page 587: ...Source Maximum Allowance Limit GROUND OPEN GROUND CLOSED Patient Lead to Ground Current Leakage Test and Patient Lead to Lead Current Leakage Test 115V 10uA 10uA 220 240V 500uA 10uA Table 10 19 Maxim...

Page 588: ...made with the Ground OPEN and CLOSED with power line polarity Normal and Reversed and with the unit OFF and ON For each combination the probe must be active to find the worst case condition NOTE Each...

Page 589: ...be into the meter EXTERNAL connector 4 Set the meter s FUNCTION switch to EXTERNAL position 5 Connect the probe for test with the connector of the console 6 Add the saline probe and the imaging area o...

Page 590: ...10 31 CAUTION When power to the unit is ON never switch the Polarity and the status of Neutral Be sure to turn OFF power to the unit before switching them using the POLARITY switch and or the NEUTRAL...

Page 591: ...lies with the probe or the scanner NOTE Each probe will have some amount of leakage dependent on its design Small variations in probe current leakages are normal from probe to probe Other variations...

Page 592: ...ormat Model Number Serial Number Probe 6 Frequency Scan Format Model Number Serial Number Probe 7 Frequency Scan Format Model Number Serial Number Probe 8 Frequency Scan Format Model Number Serial Num...

Page 593: ...Source Current Leakage CRT Patient Lead Source Leakage Lead to Ground Patient Lead Source Leakage Lead to Lead Patient Lead Source Leakage Isolation Peripheral 1 Current Leakage Peripheral 1 Ground C...

Page 594: ...mericas Fax 414 544 3384 P O Box 414 Milwaukee Wisconsin 53201 0414 U S A GE Medical Systems Europe Fax 33 1 40 93 33 33 Paris France GE Medical Systems Asia Fax 65 291 7006 Singapore GE Vingmed Ultra...

Reviews: