Chapter 7 - Diagnostics/Troubleshooting
7-25
7) Proceed as follows:
•
If no problems were identified, proceed to the FB tests, as described in the
Assembly (FB) Diagnostic Tests
•
If no problems were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
•
If errors were identified but the results were marginal, perform calibration, as described in the
section, on page 6-11, and then repeat the test.
•
If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in
. If the problem persists, contact your local GE Medical Systems field
engineer for assistance.
8) When finished, trackball to the
Cancel
button and press
Select
. The
FE Diagnostics
dialog box
is
re-displayed.