March 2004
6-4
Error Codes and Diagnostics
Displaying Diagnostic
Parameters
The Model UTX878 offers built-in
Diagnostic Parameters
to aid in
the troubleshooting of transducer and/or electrical problems. To
access these parameters, do the following:
1.
From the display screen, press [ESC].
Note:
If the display screen is locked, you will need to enter [ESC],
[ENT], [ESC] and the security code. Refer to page 2-2 for
details.
2.
Press the [
X
] key. The screen will highlight the
channel
you wish
to check. Be sure the desired channel appears on the screen (or
change it, as discussed in Chapter 3).
3.
Press the [
X
] key to scroll to the
measurement
entry in the upper
right. Press [ENT]. Be sure
Diagnostic
is highlighted, and press
[ENT].
4.
Press the [
X
] key twice to access the
units
parameter. Press
[ENT]. Scroll to the desired Diagnostic unit (as described in Table
6-1 on the next page) and press [ENT].
5.
If desired, repeat steps 1-4 for the other channel.
Summary of Contents for UTX878 Ultrasonic
Page 2: ...iii March 2004 The UTX878 Flow Transmitter and Transducers ...
Page 8: ...Chapter 1 ...
Page 20: ...March 2004 Installation 1 11 Figure 1 5 UTX878 Outline and Installation Dwg 712 1196 rev 2 ...
Page 22: ...March 2004 Installation 1 13 Figure 1 7 UTX878 Wiring Diagram Dwg 702 528 rev 2 ...
Page 23: ...Chapter 2 ...
Page 45: ...Chapter 3 ...
Page 52: ...Chapter 4 ...
Page 60: ...Chapter 5 ...
Page 65: ...Chapter 6 ...
Page 75: ...Chapter 7 ...
Page 80: ...Appendix A ...
Page 85: ...Appendix B ...
Page 86: ...CE Mark Compliance Introduction B 1 EMC Compliance B 1 ...
Page 88: ...Appendix C ...
Page 89: ...Data Records Site Data C 1 ...
Page 93: ...Appendix D ...
Page 94: ...Service Record Introduction D 1 Data Entry D 1 Diagnostic Parameters D 3 ...