March 2004
2-18
Programming Site Data
Entering the Meter
Correction (K) Factor
With the
K Factor
submenu, you can calibrate or adjust the UTX878
readings to another flow reference. While following the programming
instructions, refer to Figure A-1 on page A-1 of Appendix A,
Menu
Maps
.To enter K Factor data:
Note:
If you are in the PROG menu, proceed directly to Step 4. If you
scroll to a different option, press [ENT] twice to select that
option (once to enter and again to confirm the selection).
1.
Press [ESC]. The UTX878 enters the
User Program
.
2.
Press the [
X
] key until
PROG
is bracketed in the top left corner
and press [ENT].
3.
Use the [
S
] and [
T
] keys to scroll to the desired
Channel
and
press [ENT].
4.
Scroll to the
K Factor
submenu and press [ENT].
5.
The screen shows three options:
Off, Single
and
Table
. Scroll to
the desired option, and press [ENT].
6.
The menu varies, depending on your selection in Step 5.
•
If you select
Off
, no further choices are available.
•
If you select
Single
, the UTX878 displays the
K Factor
. To change
the value, press [ENT]. Use the arrow keys to change the value and
press [ENT].
IMPORTANT:
If you have enabled the Reynolds Correction factor
in the Fluid option, the K factor should be set to 1.00.
Otherwise, the typical factor is between 0.5 and 2.00.
•
If you select
Table
, the screen displays three options:
Units, Rows
Used
and
Edit Table
.
a.
If you scroll to
Units
and press [ENT], the screen displays the
selected measurement units (either metric or English). Press
[ESC] or [ENT] to return to the previous screen.
b.
If you select
Rows Used
, the program asks for the number of
rows
you wish to use. Enter the desired number (from 2 to 20)
and press [ENT].
c.
If you select
Edit Table
, the table opens with a series of
rows
.
Scroll to the desired row, and press [ENT].
d.
For each row, the screen displays the X (velocity) value and the
K Factor. If you wish to change either value, scroll to the value
and press [ENT]. Use the arrow keys to change the value and
press [ENT].
e.
Repeat steps c and d until you have programmed all available
data (from 2 to 20 rows).
Press [ESC] until you return to the
PROG
menu, or continue pressing
[ESC] until the display screen reappears.
Summary of Contents for UTX878 Ultrasonic
Page 2: ...iii March 2004 The UTX878 Flow Transmitter and Transducers ...
Page 8: ...Chapter 1 ...
Page 20: ...March 2004 Installation 1 11 Figure 1 5 UTX878 Outline and Installation Dwg 712 1196 rev 2 ...
Page 22: ...March 2004 Installation 1 13 Figure 1 7 UTX878 Wiring Diagram Dwg 702 528 rev 2 ...
Page 23: ...Chapter 2 ...
Page 45: ...Chapter 3 ...
Page 52: ...Chapter 4 ...
Page 60: ...Chapter 5 ...
Page 65: ...Chapter 6 ...
Page 75: ...Chapter 7 ...
Page 80: ...Appendix A ...
Page 85: ...Appendix B ...
Page 86: ...CE Mark Compliance Introduction B 1 EMC Compliance B 1 ...
Page 88: ...Appendix C ...
Page 89: ...Data Records Site Data C 1 ...
Page 93: ...Appendix D ...
Page 94: ...Service Record Introduction D 1 Data Entry D 1 Diagnostic Parameters D 3 ...