Programming Site Data
2-3
March 2004
Locking the UTX878
You can access the security level in two ways.
From the display screen:
1.
Press the [
X
] key three times, until the lock in the upper right
corner is highlighted.
2.
Press [ENT], and proceed to step 4 below.
From the User Program:
1.
Press [ESC]. The UTX878 enters the User Program.
2.
Press the [
X
] key until
USER
is bracketed.
3.
The menu highlights
Set Security
. Press [ENT].
4.
The screen shows three options:
•
Full Lock
, which prevents a user from changing any part of the
display or user program without the appropriate code:
•
Prog Lock
, which allows a user to change the display but not to
enter the user program:
•
Unlocked
, which allows access to both the display and the user
program.
Scroll to the desired option and press [ENT] twice.
5.
Press [ESC] to return to the User Program, or continue pressing
[ESC] to return to the display screen. If you have chosen to fully
lock the UTX878, the screen appears similar to Figure 2-3 below,
with a solid lock in the upper right corner. (For a meter with only
the user program locked, the lock shows a keyhole in the center.)
Figure 2-3: UTX878 Screen with Locked Program
Ch 1 Velocity
0.0
Meters/sec E1
Summary of Contents for UTX878 Ultrasonic
Page 2: ...iii March 2004 The UTX878 Flow Transmitter and Transducers ...
Page 8: ...Chapter 1 ...
Page 20: ...March 2004 Installation 1 11 Figure 1 5 UTX878 Outline and Installation Dwg 712 1196 rev 2 ...
Page 22: ...March 2004 Installation 1 13 Figure 1 7 UTX878 Wiring Diagram Dwg 702 528 rev 2 ...
Page 23: ...Chapter 2 ...
Page 45: ...Chapter 3 ...
Page 52: ...Chapter 4 ...
Page 60: ...Chapter 5 ...
Page 65: ...Chapter 6 ...
Page 75: ...Chapter 7 ...
Page 80: ...Appendix A ...
Page 85: ...Appendix B ...
Page 86: ...CE Mark Compliance Introduction B 1 EMC Compliance B 1 ...
Page 88: ...Appendix C ...
Page 89: ...Data Records Site Data C 1 ...
Page 93: ...Appendix D ...
Page 94: ...Service Record Introduction D 1 Data Entry D 1 Diagnostic Parameters D 3 ...