Chapter 2. Installation
28
Moisture Target Series 6 Hygrometer User’s Manual
2.6.4a Connecting the High/Low Alarms (A and B)
IMPORTANT:
Ensure that the power is off before proceeding.
Each of these alarms can be set to trip on either a high or low condition. For
a high alarm, the alarm will trip if the input exceeds the setpoint. For a low
alarm, the alarm will trip if the input drops below the setpoint. Make
connections to the Alarm A and Alarm B upper terminal block on the back
of the
MTS6
, as shown in Figure 25 and Figure 26 on page 19.
IMPORTANT:
To maintain good contact at each terminal block and to avoid
damaging the pins on the connector, pull the connector
straight off (not at an angle), make cable connections while the
connector is away from the unit, and push the connector
straight on (not at an angle) when the wiring is complete.
Figure 37: Removing the Upper Connector