Chapter 3. Operation and Programming
Moisture Target Series 6 Hygrometer User’s Manual
73
3.6.10 Selecting the Probe Type
Note:
Use the following procedure to select the probe type.
IMPORTANT:
Changing the probe setting will default the output and alarm
settings to DPC. Also, the default fault trip point values will be
set along with the default output range values.
From the
Settings Menu
select
Probes
and
press
Enter
. The following screen
appears.
Note:
To access the
Probes
menu, the
User
Passcode
is required (see section 3.6.1 on
Use the left and right arrow keys to select
each digit to be changed and the up and
down arrow keys to increase or decrease its
value. Once the passcode has been entered,
press
Enter
and the following screen
appears.
Use the left or right arrow key to select the
correct probe type and press
Enter
.
The
MTS6
reboots in 5 seconds.