Chapter 4. Service and Maintenance
82
Moisture Target Series 6 Hygrometer User’s Manual
Table 3: Troubleshooting Guide for Common Problems
Possible Cause
Response and Action
Symptom: The accuracy of the moisture sensor is questioned.
There is insufficient
time for the system to
equilibrate.
Response:
Reads too wet during dry down conditions
or too dry in wet up conditions.
Action:
Change the flow rate. A change in dew point
indicates the sample system is not at equilibrium or
there is a leak. Allow sufficient time for sample system
to equilibrate and moisture reading to become steady.
Check for leaks.
Dew point at the
sampling point is
different from the dew
point of the main
stream.
Response:
Reads too wet or too dry.
Action:
Readings may be correct if the sampling point
and main stream do not run under the same process
conditions. The different process conditions cause
readings to vary. If sampling point and main stream
conditions are the same, check sample system pipes,
and any pipe between the sample system and main
stream for leaks. Also, check sample system for
adsorbing water surfaces, such as rubber or plastic
tubing, paper-type filters, or condensed water traps.
Remove or replace the contaminating parts with
stainless steel parts.
Sensor or sensor
shield is affected by
process contaminants
Response:
Reads too wet or too dry
Action:
Clean the sensor and the sensor shield, then
reinstall the sensor.
Sensor is
contaminated with
conductive particles.
Response:
Reads high dew point.
Action:
Clean the sensor and the sensor shield, then
reinstall the sensor. Also, install a proper filter (i.e.
sintered or coalescing element).
Sensor is corroded
Response:
Reads too wet or too dry
Action:
Return probe to factory for evaluation.
Stream particles
causing abrasion.
Response:
Reads too wet or too dry.
Action:
Return probe to factory for evaluation.
Summary of Contents for Moisture Target Series 6
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 10: ...Contents viii Moisture Target Series 6 Hygrometer User s Manual ...
Page 90: ...Chapter 3 Operation and Programming 80 Moisture Target Series 6 Hygrometer User s Manual ...
Page 126: ...Index 116 Moisture Target Series 6 User s Manual ...
Page 130: ......
Page 131: ......