Work with the Files List
To print a copy
of the patient
report ...
Perform the following:
In a different
report format
1.
Select anywhere around the ellipsis in the left, bottom corner of the
screen.
2.
From the expanded
Copy Format
menu, select the report format.
For example, if a 12-lead ECG is recorded, you can only select 12-lead
ECG patient report formats.
3.
Select the
Copy
icon:
The patient report is refreshed and displayed on the
ECG
or
FD Report
or
Rhythm
tab in the selected report format. A job to print one copy of the
ECG or rhythm in the selected report format is sent to the printer.
The patient report is printed in the order in which it was received. If no other patient
reports are printing, the report is printed immediately. You will see a progress
message at the bottom of the screen indicating the printing status.
NOTE
:
Based on the
Mandatory fields apply for
Acquisition
settings, the ECG report
is not accepted, transmitted, or printed until you enter the patient demographic
data for the mandatory fields. If you manually print, an error message
Unable to
print. Incomplete patient data.
displays on the
Acquisition
screen. You need to
complete the data for the mandatory fields to print the patient report.
If a printer error occurs, the progress message is replaced by the printer error. The
printing restarts automatically after the error is resolved. For more information on
printer errors, see
.
If the device is configured to print the barcode of the patient ID in the patient reports,
the printed patient report includes the barcode. The barcode can be used to perform
a query by patient ID in the MUSE system.
To stop printing a patient report, do one of the following:
• Press the
Stop
key on the keyboard:
• Press the
Stop
button on the Acquisition Module:
Print a List of Stored Records
You can print all the stored records that display in the
Files Manager
.
2088531-370-2
MAC VU360
™
Resting ECG Analysis System
93