Recording an ECG or Rhythm
Item
Name
Description
7
Copy
icon
Prints a copy of the patient report. For more information, see
.
8
Transmit
icon
Transmits the patient report. For more information, see
Patient Report to a Configured Destination on page 60
9
Live
tab
Select the
Live
tab to return to the live waveform display in the
Acquisition screen.
10
Full Disclosure
tab
Click the
Full Disclosure
tab from the patient record menu bar to
navigate to the Full Disclosure waveform and application.
NOTE
:
The
Full Disclosure
tab displays after you purchase and enable
the Full Disclosure option.
To start a new ECG on the same patient, select the
Live
tab to return to the
live waveform display in the Acquisition screen, and restart the ECG. For more
information, see
ECG Acquisition Overview on page 47
.
After you select
Done
on the
Review
screen of an ECG patient report, disconnect
patient leads before you select
Start New Patient
on the MAC VU360 device.
To start an ECG for a new patient, select
Start New Patient
. For more information,
see
Start a Test for a New Patient on page 31
.
To start a new 10 seconds ECG on the same patient from the Full Disclosure tab,
click the
Full Disclosure
tab from the patient record menu bar to navigate to the
Full Disclosure waveform and application, and select 10 seconds ECG from the Full
Disclosure waveform. For more information, see
Record a Full Disclosure ECG on page
.
Automatically Print an ECG Patient Report
When a patient report is saved in the
Files
list, it is automatically printed in the
configured report formats.
The configured report formats determine the following:
• Number of copies printed
• Inclusion or exclusion of 12SL interpretation statements
• Printing of all reports
• Printing of only the reports interpreted by the 12SL analysis as abnormal
The ECG patient report is printed in the order in which it was received. If no other
patient reports are printing, the report is printed immediately.
You will see a progress message at the bottom of the screen indicating the printing
status.
56
MAC VU360
™
Resting ECG Analysis System
2088531-370-2
Summary of Contents for MAC VU360