Report Formats
Item
Description
8
Prints the product name.
9
Prints the
Filter
of the ECG waveform (measured in Hz), indicated with ZPD (for High Pass
filter).
NOTE
:
You can change the filter after recording and/or printing of a rhythm.
For the printed FD report, the recording stops and restarts at the newly selected
filter. Each time the Full Disclosure recording starts after a change in filter, a
calibration pulse is added for each lead that tells which filter on the printed rhythm.
For the digital FD report, the calibration pulse of the last filter selected during
recording is shown on the full report.
10
Prints the
Gain
of the ECG waveform (measured in mm/mV).
NOTE
:
You can change the gain after recording and/or printing of a rhythm.
For the printed FD report, the recording stops and restarts at the newly selected
gain. Each time the Full Disclosure recording starts after a change in gain, a
calibration pulse is added for a single lead that tells the gain on the printed FD
report.
For the digital FD report, the calibration pulse of the last gain selected during
recording is shown on the full report.
11
Prints the
Speed
of the ECG waveform (measured in mm/s).
NOTE
:
The digital FD report is recorded at the configured rhythm speed.
12
Displays the time stamp for each row.
The time stamp corresponds to clock time when the first sample is acquired in that row.
13
Prints the
Calibration pulses
.
When a rhythm recording is started, a calibration pulse is added at the beginning of a
single lead in the patient report , which shows the speed and gain at which the rhythm
is recorded. Each calibration pulse represents 1 mV of amplitude and 200 ms duration of
the waveform. Each time the Full Disclosure recording starts after a change in speed or
gain, a calibration pulse is printed for a single lead.
NOTE
:
The standard grid paper is divided into small squares of 1 mm x 1 mm and large
squares of 5 mm x 5 mm. When recording 25 mm/s, 10 rows of data, 10 seconds of
data per row is shown in each page on the FD report.
142
MAC VU360
™
Resting ECG Analysis System
2088531-370-2