Login and Security
NOTE
:
The
Restore to Factory Defaults
resets the settings or section of settings.
Use this procedure as a last solution. Transfer your data from the system before you
start the procedure.
1.
When the
Login
screen displays, press and hold
Shift + Esc
for 5 seconds.
The
System Restore
screen displays a warning that the System Restore will
return your system to the original factory shipped configuration. All patient data,
system setup changes, logs, and user data will be lost and unrecoverable.
2.
Enter the serial number of the device in the
Enter the system serial number
field. The
Restore
button enables when the serial number is entered correctly.
3.
Select
Restore
to proceed with system restore.
The system configuration is reset to factory defaults and all patient data records
are deleted. The device reboots. You can access the device as the
Default
user
without login credentials.
4.
To reconfigure the device, access the
Settings
screen from the user menu. A
login screen opens. Log on as
Admin
user with the default password
admin123
.
28
MAC VU360
™
Resting ECG Analysis System
2088531-370-2
Summary of Contents for MAC VU360
Page 7: ...Glossary 167...