Figure 2-30 Fluoro UPS CE dimensions (Optional)
The UPS cabinet can be positioned against the wall but, in order to improve the ventilation and to make
easier the maintenance operations for UPS and battery, we recommend a minimum distance of 200 mm
(8”) from the wall.
For maintenance operations, a minimum clearance distance of 500 mm (20”) is required for both left and
right sides of the UPS cabinet.
In front of the cabinet, the clear width of the service area to insure electrical safety shall be at least 0.9
m. In cases where 2 cabinets are installed face to face (both sides of the access way), the clear width
shall be at least 1.2 m.
Recommended minimum clearance between ceiling and top of the UPS should be 400 mm (16”) for
proper cooling air exhaust.
NOTE
A minimum distance of 2 m (79 in) between X-Ray tube Chiller and Fluoro UPS CE cabinet
is required.
NOTE
A Fire extinguisher (non-water type, ex. CO2) shall be provided and installed by the
customer close to the Fluoro UPS CE cabinet.
Equipment Requirements
66
Innova
TM
IGS 6
5750182-1EN 3
Summary of Contents for Innova IGS 6
Page 9: ...Page intentionally left blank...
Page 57: ...Figure 2 11 Omega V Table dimensions Equipment Requirements 46 InnovaTM IGS 6 5750182 1EN 3...
Page 61: ...Figure 2 14 Gas box outlets Omega Table Equipment Requirements 50 InnovaTM IGS 6 5750182 1EN 3...
Page 69: ...Figure 2 22 C2 Cabinet dimensions Equipment Requirements 58 InnovaTM IGS 6 5750182 1EN 3...
Page 147: ...Page intentionally left blank Environmental Requirements 136 InnovaTM IGS 6 5750182 1EN 3...
Page 158: ...Figure 5 6 PDB Schematic CE 1 2 Electrical Requirements 5750182 1EN 3 InnovaTM IGS 6 147...
Page 159: ...Figure 5 7 PDB Schematic CE 2 2 Electrical Requirements 148 InnovaTM IGS 6 5750182 1EN 3...
Page 161: ...Figure 5 9 PDB Schematic UL 1 2 Electrical Requirements 150 InnovaTM IGS 6 5750182 1EN 3...
Page 194: ...InnovaTM IGS 6...